The delay period for an infusion can be programmed as a
specific number of minutes or a time of day, as described in
the following procedures. An infusion delay can be
programmed prior to or after an infusion is initiated.
Delay Options can be enabled at the time the Alaris
®
System
is configured for use. If Delay Options is enabled, a primary
infusion can be programmed to be delayed for a specified
period of time and a callback can be scheduled, as described
in the following procedures.
Since by definition, an infusion with Delay Options will not be
infusing for a programmed period of time, it is assumed that
another infusing
IV
line will keep the vein open until the
delayed infusion begins. When a delay is programmed, the
infusion stops when complete and no
KVO
is delivered.
The
Delay for
option is used to program an infusion delay for
a minimum of
1
minute and up to
120
minutes.
1.
Press
DELAY OPTIONS
soft key.
Programming
2-57
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Specifying by Minutes
Delaying Infusion
>Press START
40
mL/h
A
250
mL
Infusion Setup
RATE
SECOND-
ARY
START
DELAY
OPTIONS
VTBI
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
Anesthesia Mode
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Reference the PC Unit Section of this
DFU
.
Delay Options
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
-- Continued on Next Page --
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......