70-00969-01 F
Powerheart® AED G3 Pro 9300P
3-14
How to Perform a Rescue
3.
Plug the ECG patient cable into the electrode socket located on
the AED.
Once connected the AED displays and evaluates the patients ECG
(Lead II) and heart rate. Follow all prompts from the AED.
Check the patient if:
◆
Indicated by the observed ECG or heart rate display.
◆
The patient becomes unresponsive or stops breathing.
◆
The AED prompts “Shockable Rhythm. Attach Defibrillation
Pads”.
If appropriate, disconnect the ECG patient cable from the AED and
connect the defibrillation pads to pads socket. Then, place the
defibrillation pads on the patient.
Note:
If the Manual Mode button is pressed when in ECG Monitoring
Mode, a voice prompt will indicate “ECG Monitoring Mode”.