70-01047-01 B
Product Safety and Setup
36
Acquisition
Lead selection
I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 and Alternate
Chest Lead (chest lead selection V2R through V9R, V7, V8,
and V9)
Modes
Automatic, automatic rhythm, or manual rhythm
Frequency
response
Meets or exceeds IEC 60601-2-51 standards
Input impedance
Meets or exceeds IEC 60601-2-51 standards
Electrode offset
tolerance
±300 mV
Sampling
8000 Samples/Sec. 2.5μV LSB, Burst acquisition <100 μS
Artifact filter
response
40 Hz, -3db
Storage
500 samples/second, 2.5μV LSB
Pacemaker display
capability
Meets or exceeds IEC 60601-2-51
Interpretation
(optional)
Diagnosis, measurements, reasons statements based on five
demographic criteria
Environmental (Electrodes)
Operating and
storage
Follow the electrode package labeling for operating and
storage. To prevent electrodes from drying out Cardiac
Science recommends storing electrodes in a reclosable
pouch.