MTF 9/15 

 

 

I

NSTALLATION

 

2.1  Unpacking & Setting Up 

 

Unpack the furnace carefully and remove any packing from inside the tube. 

 

Place the furnace in a well ventilated area, away from other sources of heat, on a surface which is 
resistant to accidental spillage of hot materials.  Do not use the furnace on an inflammable surface. 

 

Ensure that there is free space around the furnace.  Do not obstruct any of the vents: they are 
needed to keep the controls cool. 

 

Ensure that the furnace is placed in such a way that it can be quickly switched off or disconnected 
from the electrical supply - see below. 

2.2  Electrical Connections 

 

The furnace requires a single-phase A.C. supply with earth (ground).  The supply may be Live to 
Neutral non-reversible, Live to Neutral with reversible plug, or Live to Live. 

 

The voltage or range of voltages on which the furnace may be operated is given on the furnace 
rating label.  Check that the supply voltage is compatible with the voltage on the label. 

 

The supply should be fused at not more than 10 Amps, preferably 2 Amps. 

 

The furnace is fitted with a supply cable which may be wired directly to an isolator which operates 
on both conductors, or fitted with a line plug which may be quickly disconnected from the supply. 

 

The supply MUST incorporate an earth (ground). 

 

Connect the supply as follows: 

 

supply type

        

 

Cable colour 

Live-Neutral 

Reversible or Live-Live 

Brown 

to live 

to either power conductor 

Blue 

to neutral 

to the other power conductor 

Green/Yellow 

to earth 

to earth 

 

3.0 

MF09 

Summary of Contents for MTF 15

Page 1: ...supply CONTENTS section page 1 0 Symbols Warnings 2 2 0 Installation 3 3 0 Operating Instructions 4 4 0 Controller Operation 5 5 0 Maintenance 10 6 0 Repairs Replacements 11 7 0 Fault Analysis 13 8 0...

Page 2: ...s Heat Light the adjacent light glows or flashes to indicate that power is being supplied to the elements 1 2 Warning Symbols DANGER of electrical shock read any warning printed by this symbol DANGER...

Page 3: ...A C supply with earth ground The supply may be Live to Neutral non reversible Live to Neutral with reversible plug or Live to Live The voltage or range of voltages on which the furnace may be operate...

Page 4: ...e at high temperature when not required The maximum temperature is shown on the furnace rating label and on the back page of this manual Materials such as waxes case hardening compounds melting fluxes...

Page 5: ...in the Navigation Diagram To return to the Home list at any time press Page and Scroll 4 together or wait for 45 seconds The Down T and Up S keys are used to alter the setpoint or other parameter valu...

Page 6: ...nd fig 2 indicate the possible difference between running without and with a ramp to setpoint value depending on the load and the value used 4 6 Navigation Diagram OP Hi SPrr m A w SP Home List Input...

Page 7: ...The timer starts timing immediately At the end of the timing period control switches off i e goes into Manual to allow cooling and End flashes on the display On during the timing period Off from the...

Page 8: ...ameter changes to run Note that the tmr display shows 0 zero during the last minute of timing and also shows 0 when the time has expired The timer light indicates whether timing is still in progress M...

Page 9: ...rrors or other system errors User calibration allows compensation for such errors and the 2132 allows for a user 2 point calibration This setting is password protected to avoid accidental alteration P...

Page 10: ...hroughout the world We also sell spares by mail order A telephone call or fax often enables a fault to be diagnosed and the necessary spare part despatched Each furnace has its own record card at Carb...

Page 11: ...ulletin No 11 and the UK Health and Safety Executive Guidance Note EH46 We can provide further information on request Alternatively our service division can quote for any repairs to be carried out at...

Page 12: ...beads When refixing the wrapped element position the beads just above edge of the metal channel to maximise the clearance between the outer case and the wires 6 6 Solid State Relay Replacement The so...

Page 13: ...ghts glowing on the controller The SUPPLY light is ON The controller may be faulty or not receiving a supply due to a faulty switch or a wiring fault The SUPPLY light is OFF Check the supply fuses and...

Page 14: ...MTF 9 15 CIRCUIT DIAGRAM 8 1 Circuit Diagram heating element thermocouple 2A 2A Re d Re d Blue Blue L N PE controller L N filter SSR pink grey 2 1 3 4 Green Yellow 9 0 14 MF09...

Page 15: ...nd 220 240V models Electronic power limiting in the 2132 controller is used to reduce the power in the 220 240V model The correct power limit settings parameter OP Hi for the furnace are as follows 22...