ELF (B) 

 

 

F

AULT 

A

NALYSIS

 

A.

 

Furnace Does Not Heat Up

 

 

 

 

 

 

1. 

The 

HEAT

 light 

 is 

ON

 

Æ

  The heating 

element has failed 

Æ

Check also that the SSR is working 
correctly 
 

2. 

The 

HEAT

 light  

is 

OFF

 

 The 

controller 

shows 

a very high 

temperature

 or a 

code such as S.br 
 

Æ

The thermocouple has broken or has a 
wiring fault 

  

 The 

controller 

shows a

 low 

temperature

 

Æ

The door switch(es) (if fitted) may be 
faulty or need adjustment 

 

 

 

 

Æ

The contactor (if fitted) may be faulty 
 

 

 

 

 

Æ

The SSR could be failing to switch on due 
to internal failure, faulty logic wiring from 
the controller, or faulty controller 
 

 

 

 

There are no 
lights glowing 
on the 
controller 

Æ

The 

SUPPLY

 

light is 

ON

 

Æ

The controller may be 
faulty or not receiving a 
supply due to a faulty 
switch or a wiring fault 
 

 

 

 

 

Æ

The 

SUPPLY

 

light is 

OFF

 

Æ

Check the supply fuses and 
any fuses in the furnace 
control compartment 
 

B.

 

Furnace Overheats

 

 

 

 

 

 

1. 

The 

HEAT

 light 

goes 

OFF

 with 

the instrument 
switch 

Æ

  The controller 

shows a 

very high 

temperature 

Æ

The controller is faulty 

 

 

Æ

  The controller 

shows a 

low 

temperature 

Æ

The thermocouple may have been shorted 
out or may have been moved out of the 
heating chamber 
 

 

 

 

 

Æ

The thermocouple may be mounted the 
wrong way round 
 

 

 

 

 

Æ

The controller may be faulty 
 

2. 

The 

HEAT

 light 

does not go off

 

with the 
instrument 
switch 

Æ

  The SSR has failed 

“ON” 

Æ

Check for an accidental wiring fault that 
could have overloaded the SSR 

 
 

9.0 

MF01 

19 

Summary of Contents for ELF 11/14B

Page 1: ...ty supply CONTENTS Section page 1 0 Introduction 1 2 0 Installation 2 3 0 Operation 5 4 0 Controller Operation 7 5 0 Overtemperature Control Option 13 6 0 Maintenance 14 7 0 Repairs Replacements 14 8...

Page 2: ...ns The B versions may be set to operate on any single phase voltage in the range 200 240V or 100 120V to order by correct setting of the power limit parameter in the controller Because of model change...

Page 3: ...pply 2 3 Chimney The chimney is a short length of ceramic tubing If it is supplied unfitted then fit it through the hole in the top of the furnace case If the furnace is to be used to heat substances...

Page 4: ...the supply cable directly to an isolator or fit it with a line plug The plug or isolator should be within easy reach of the operator to permit quick disconnection of the power For operator safety the...

Page 5: ...controller setpoint to a temperature close to the desired maximum Materials such as case hardening compounds and other reactive salts may attack the wire elements causing premature failure The insula...

Page 6: ...ELF B 4 0 6 MF01...

Page 7: ...oint ramp rate is Off A single press of the scroll key 4 displays the temperature units further presses reveal the parameters in the current list indicated in the Navigation Diagram To return to the H...

Page 8: ...ate the possible difference between running without and with a ramp to setpoint value depending on the load and the value used 4 6 Navigation Diagram OP Hi SPrr m A w SP Home List Input List Output Li...

Page 9: ...he timer starts timing immediately At the end of the timing period control switches off i e goes into Manual to allow cooling and End flashes on the display On during the timing period Off from the en...

Page 10: ...r changes to run Note that the tmr display shows 0 zero during the last minute of timing and also shows 0 when the time has expired The timer light indicates whether timing is still in progress Method...

Page 11: ...rs or other system errors User calibration allows compensation for such errors and the 2132 allows for a user 2 point calibration This setting is password protected to avoid accidental alteration Page...

Page 12: ...once to display the value of OP Hi it is good practice to write down the original value To alter the parameter to the desired value use Down T or Up S Caution Do not increase the power limit value to...

Page 13: ...ons the letters are described as m and w With the key pressed the overtemperature setpoint is normally displayed To alter the value use Down and Up To view the measured temperature press Scroll until...

Page 14: ...on Thermal Engineering Services has a team of Service Engineers capable of repair calibration and preventive maintenance of furnace and oven products at our customers premises throughout the world We...

Page 15: ...nt Wash work clothing separately Before commencing any major repairs we recommend reference to the European Ceramic Fibre Industry Association Bulletin No 11 and the UK Health and Safety Executive Gui...

Page 16: ...a socket spanner undo the four bolts that fasten the insulation box These are located under the insulation box in the two cross supports Remove the insulation box by carefully sliding it out through...

Page 17: ...rder DOOR DOOR PLUG CARRIER DOOR PLUG 7 8 Air Inlet Adjustment The air inlet is on the inner door panel door plug carrier and is supplied in the closed position It can be altered to the open position...

Page 18: ...visable for an electrician to check the internal circuits Replace any failed fuses with the correct type For safety reasons do not fit larger capacity fuses without first consulting Carbolite The fuse...

Page 19: ...hts glowing on the controller The SUPPLY light is ON The controller may be faulty or not receiving a supply due to a faulty switch or a wiring fault The SUPPLY light is OFF Check the supply fuses and...

Page 20: ...n t thermocouple EMC filters circuit boards 9 3 Fuses F1 F3 Refer to the circuit diagram F1 Internal supply fuses 32mm x 6mm type F fitted on EMC filter circuit board s F2 Auxiliary circuit fuses 20mm...

Reviews: