background image

 
1700-1800° Tube 

 

 

M

AINTENANCE

 

4.1  Routine Maintenance 

 

No routine maintenance is required. The outer surfaces may be cleaned with a damp cloth. Do not 
allow water to enter the interior of the case, tube or control box. Do not clean with organic 
solvents. 

4.2 

Calibration 

 

After prolonged use the controller and/or thermocouple could require recalibration; see the 
warning about thermocouples in section 3.8. This would be important for processes which require 
accurate temperature readings or which use the furnace close to its maximum temperature. A 
quick check using an independent thermocouple and temperature indicator should be made from 
time to time to determine whether full calibration is required. Carbolite can supply these items. 

 

Depending on the controller, the controller manual may contain calibration instructions. 

4.3 

After Sales Service 

Carbolite’s service division (Thermal Engineering Services) has a team of Service Engineers 
capable of repair, calibration and preventive maintenance of furnace and oven products at our 
customers’ premises throughout the world. We also sell spares by mail order. A telephone call or 
fax often enables a fault to be diagnosed and the necessary spare part despatched.  

 

Each furnace has its own record card at Carbolite. In all correspondence please quote the serial 
number, model type and voltage given on the rating label of the furnace. The serial number and 
model type are also given on the front of this booklet when supplied with a furnace.  

 

To contact Thermal Engineering Services or Carbolite see the back page of this manual. 

4.4  Recommended Spares Kits 

 

Carbolite can supply individual spares, or a kit of the items most likely to be required. Ordering a 
kit in advance can save time in the event of a breakdown. Each kit comprises one thermocouple, 
one sheath, one power thyristor, and a set of elements, clips and braids. It is advisable also to 
obtain element clamps and insulators (not included in kit). 

 

When ordering spares please quote the model details as requested above. 

4.5 

Power Adjustment (Controller) 

 

The furnace controller incorporates a power limit parameter 

OP.Hi

 which is usually inaccessible to 

the operator. 

 

Occasionally the power limit is set to zero to permit demonstration of the controls without the 
heating elements taking power. In this case the power limit is accessible to the operator and may 
be reset to its standard value of 100. 

4.6  Power Adjustment (Thryristor) 

The current-limiting thyristor stacks which control power to the elements are fitted with an 
adjustable resistor which is factory set to limit the maximum current supplied. In the event of a 
change of supply voltage, or the fitting of a new thyristor, further adjustment may be required.  

 

The maximum element currents for the various models are listed in section 8.2. Please contact 
Carbolite for further information. 

4.7  Work Tube Supports 

Occasionally the internal work tube supports may need adjusting. See section 2.8. 

5.0 

10 

MF32 

Summary of Contents for CTF

Page 1: ...d to the electricity supply CONTENTS page section 1 0 Symbols Warnings 1 2 0 Installation 3 3 0 Operation 6 4 0 Maintenance 10 5 0 Repairs Replacements 11 6 0 Fault Analysis 14 7 0 Circuit Diagram 14 8 0 Fuses Power Control 16 9 0 Specifications 17 Manuals for the furnace controller and overtemperature controller are supplied separately Please read the controller manuals before operating the furna...

Page 2: ... in the adjacent switch glows Heat Light the adjacent light glows or flashes to indicate that power is being supplied to the elements 1 2 Warning Symbols DANGER of electrical shock read any warning printed by this symbol DANGER hot surface Read any warning printed by this symbol WARNING all surfaces of a furnace may be hot 2 0 2 MF32 ...

Page 3: ... material therein Carefully remove all packing material Note that in the TZF models there are zone barriers comprising insulation pieces which protrude into the chamber take care not to damage these pieces Model TZF 18 600 see separate booklet MS12 Reassemble the insulation by reversing the removal procedure Now fit the heating elements 2 4 Fitting the Heating Elements Wear eye protection when han...

Page 4: ...l label Cable colour Live Neutral Reversible or Live Live 1 phase L Brown To live to either power conductor N Blue To neutral to the other power conductor PE Green Yellow To earth ground to earth ground Supply Terminal label Cable colour 3 phase L1 Black to phase 1 L2 Black to phase 2 L3 Black to phase 3 N Light Blue to neutral except delta PE Green Yellow to earth ground 2 7 Inserting a Tube If a...

Page 5: ...er the support tubes if the work tube is removed or replaced and later readjust them 2 9 Tube Fittings if ordered For optimum temperature uniformity insulating plugs should be placed in the tube ends as shown in fig 1 With a long work tube the stem of the plug assembly should line up with the end of the tube as in fig 2 Alignment of radiation shields is similar to that of plugs If stainless steel ...

Page 6: ...eft off isolate it from the electrical supply 3 2 Operating Cycle TZF See section 3 3 for a description of the control methods A B C The furnace is fitted with a combined Supply light and Instrument switch The light is on whenever the furnace is connected to the supply The switch cuts off power to the controllers and contactor Connect the furnace to the electrical supply The Supply light should gl...

Page 7: ... this type of control does not cool down the end controllers try to remain at the central temperature Should this occur contact Carbolite s technical department 2132 Slave Controller Navigation Diagram Home List Access List C F k ACCS C F k codE 4 Zone temperature difference home list 0 0 Used in factory configuration not accessible to the operator B Retransmission of Setpoint In this system the c...

Page 8: ...ot arise Poor thermal contact should be encouraged between the workpiece and the tube crucibles or boats should be of low thermal mass and should have feet to reduce the contact with the tube fig 4 Avoid metal directly touching the work tube see the safety warning above Do not set too high a heating rate Large diameter tubes are more susceptible to thermal shock than smaller A general rule for max...

Page 9: ...1700 1800 Tube 4 0 MF32 9 ...

Page 10: ...he serial number and model type are also given on the front of this booklet when supplied with a furnace To contact Thermal Engineering Services or Carbolite see the back page of this manual 4 4 Recommended Spares Kits Carbolite can supply individual spares or a kit of the items most likely to be required Ordering a kit in advance can save time in the event of a breakdown Each kit comprises one th...

Page 11: ... repairs we recommend reference to the European Ceramic Fibre Industry Association Bulletin No 11 and the UK Health and Safety Executive Guidance Note EH46 We can provide further information on request Alternatively our service division can quote for any repairs to be carried out at your premises or ours 5 3 Temperature Controller Replacement 2132 2416 2408 etc Ease apart the two lugs at the side ...

Page 12: ...he element diagram shows the items which must be assembled and fitted Examine and identify the loose components There are variations in the positions of the braids which connect the elements the rule is that for single phase all elements are in series for 3 phase or 3 zone each phase or zone has elements in series Fit the insulation piece element block to the element Locate the clamps over the ele...

Page 13: ...17 600 Example CTF 17 300 Note 300 models have 6 elements in series CTF 600 single phase models have 10 elements in series CTF 600 3 phase models have 4 els in S 2 els in S 4 els in S TZF 600 models have 2 els ins S 6 els in S 2 els in S 6 0 MF32 13 ...

Page 14: ...e controller may be faulty or not receiving a supply due to a faulty switch or a wiring fault Æ The SUPPLY light is OFF Æ Check the supply fuses and any fuses in the furnace control compartment B Furnace Overheats 1 The HEAT light goes OFF with the instrument switch Æ The controller shows a very high temperature Æ The controller is faulty Æ The controller shows a low temperature Æ The thermocouple...

Page 15: ...AGRAM 7 1 Example CTF 17 300 single phase 7 2 Example TZF 17 600 Three phase Neutral back to back thermocouples TC temperature programmer OTC overtemperature controller GR Y Green Yellow R Red BU Blue P Pink G Grey 8 0 MF32 15 ...

Page 16: ...Model Phases volts supply fuse rating type thyristor fuse rating Current limit element circuit CTF 17 300 1 phase 220 240 30 32A NS32 50A 90A CTF 17 600 1 phase 220 240 50A ES50 63A 90A CTF 17 600 3 phase 380 220 415 240 25A NS25 2 x 30A 1 x 20A 134A CTF 17 600 3 phase delta 220 240 25A NS25 2 x 30A 1 x 20A 134A TZF 17 600 1 phase 220 240 50A ES50 2 x 20A 1 x 50A 90A TZF 17 600 3 phase 380 220 415...

Page 17: ...ng the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specifications are subject to change without notice The user is responsible for determining the suitability of this product ...