Quick Start Guide
8
MC19-1.22
4 Setting up the controller
Before using the controller (or during its lifetime) certain parameters may have to be set,
depending on specific requirements. To do this the Controller must be set to supervisor
Level 2, see section 3.5.
4.1
Maximum output power setting
Press page
until Control Output Hi is displayed. Press up
or down
to adjust the value.
Depending on the furnace or oven model, the maximum power output setting may be
accessible or locked.
For Silicon Carbide heated furnaces the parameter is accessible to allow compensation for
element ageing. Refer to the product manual for details.
In many models the maximum power output setting depends on the supply voltage, refer to
the product manual for details.
4.2
Customer ID.
A furnace or oven identity number can be entered if required. This may be used to identify
one of many units, for production or quality control systems.
Press page
until Customer Identity is displayed. Press up
or down
to select a number.
4.3
Units
Press page
until Units is displayed. Press up
or down
to select:
C
Celsius
F
Fahrenheit
K
Kelvin
4.4
Language
The text on the 3508 can be shown in different languages, this can only be set at the factory
and therefore must be specified at the time of placing an order.