2216 Controller

 

 

 
MC11 – 1.05 

2

 

2216

 

 

B

ASIC 

O

PERATING 

I

NSTRUCTIONS

 

2.1

 

Furnace or Oven Controls 

 

Most Carbolite Gero products are fitted with an “Instrument Switch” which cuts off power to the 
controller and other parts of the control circuit. See the instruction manual for the furnace or oven 
for the overall operating instructions. 

 

To operate the 2216 there must be power to the furnace or oven, and the Instrument switch must 
be on. If a time switch is included in the furnace or oven circuit, this must be in an On period. 

2.2

 

The 2216 – Operation 

 

 

When switched on, the controller lights up, goes through a short test routine, and then displays 
the measured temperature and starts to control. The output light glows or flashes as heating 
occurs.  

 

The 

Page

 key 

 allows access to parameter lists within the controller; most lists and parameters 

are hidden and cannot be accessed by the operator because they contain factory-set parameters 
which should not be changed.  

 

A single press of the page key 

 displays the temperature units, normally set to °C; further 

presses reveal the lists indicated in the Navigation Diagram in section 2.6.  The 

Scroll

 key 

 

allows access to the parameters within a list. Some parameters are display-only; others may be 
altered by the operator. Some parameters only appear in appropriate circumstances – for 
example, working setpoint does not appear if setpoint ramp rate is Off.  

 

A single press of the scroll key 

 displays the temperature units; further presses reveal the 

parameters in the current list indicated in the Navigation Diagram.  

 

To return to the Home list at any time, press Page 

 and Scroll 

 together, or wait for 45 

seconds. 

 

The 

Down

 

 and 

Up

 

 keys are used to alter the setpoint or other parameter values. 

2.3

 

Basic Operation 

Normally no operator action is required other than entering the setpoint, as the 2216 starts to 
control on being switched on, as described above. 

Summary of Contents for Eurotherm 2216

Page 1: ...16 Overtemperature Controller Instructions See also the main manual for the oven or other product to which the controller is fitted CONTENTS section page 1 2216 Description 2 2 2216 Basic Operating In...

Page 2: ...d as a sleeve fitted controller The 2216 is supplied configured for easy use as a simple temperature controller where on setting the required temperature the controller immediately attempts to reach a...

Page 3: ...den and cannot be accessed by the operator because they contain factory set parameters which should not be changed A single press of the page key displays the temperature units normally set to C furth...

Page 4: ...te of heating or cooling in degrees per minute A value of OFF cancels the ramp rate allowing heating and cooling at the maximum rate When this feature is in use there is a working setpoint which can b...

Page 5: ...or Up once to display the value of OP Hi and write down the value To alter the value use Down or Up Note that setting the value to zero prevents the furnace or oven from heating Caution Do not increas...

Page 6: ...the controller is reset Other components do not generally have power to them cut on overtemperature oven fans usually remain running but may not do so if other options such as a door switch are fitted...

Page 7: ...ition is cleared by the temperature falling when normal operation resumes If the alarm is acknowledged when the temperature has dropped or after the overtemperature setpoint has been raised such that...

Page 8: ...Telephone 44 0 1433 620011 Fax 44 0 1433 621198 E mail info carbolite gero com For preventive maintenance repair and calibration of all Furnace and Oven products please contact Carbolite Gero Service...

Reviews: