CAPINTEC, INC
CRC
®
-25R
Place Cartridge
in Chamber
Any Key to Continue
Figure 11-57 MAG3 Measurement Message – Cartridge
Place the Cartridge in the Chamber and press any key (except
HOME
) to perform the
measurement. Figure 11-55 QC Test Measurement Screen will re-appear.
Press
ENTER
to accept the measurement. Figure 11-58 MAG3 Results Screen will appear.
QC TEST MAG3
Non-elutable Tc99m
xx.x%
Hydrophilic Impure
xx.x%
Tc99m Mertiatide
xx.x%
Any Key to Continue
Figure 11-58 MAG3 Results Screen
Press any key (except
HOME
) to continue. If a printer is attached to the system, a QC Test
MAG3 report will print. The report will show the measured values and results.
The prompt to print again will appear. To print the test results again, press
YES
. Press
NO
to
continue. Figure 11-59 Perform Another Test Screen will appear.
Do Another
Test? Y or N
Figure 11-59 Perform Another Test Screen
To perform another test, press
YES
. Figure 11-54 MAG3 Measurement Message – Fraction
#1 will re-appear.
If another test is not desired, press
NO
to exit. Figure 11-1 Tests Menu will appear.
11 - 24
ENHANCED TESTS
August 07
Summary of Contents for CRC-25R
Page 2: ......
Page 12: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 26: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 58: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 104: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 122: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 192: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 208: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 226: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 227: ......
Page 228: ......