Service Manual –
endoSTORE® neo & one-click
page 9 of
28
Version 01
Revision 06
2.5.3.1
Displaying of the adjusted pressure threshold
1. Press
displaying of the current pressure.
2. Press
2
nd
time
“P1L” appears (lower threshold)
3. Press
3
rd
time
“P1H” appears (upper threshold)
4. Press
4
th
time
current pressure is displayed
2.5.3.2
Adjust Pressure of threshold
1. Press
current pressure is displayed
2. Press
2
nd
time
“P1L” appears
3. Adjust lower threshold with
and
,
or continue to step 4
4. Press
3
rd
time
“P1H” appears
5. Adjust upper threshold with
and
,
or finish with Step 6
6. Press
4
th
time
current pressure is displayed
2.5.3.3
Completing adjustment of the pressure regulator
In case that the pressure regulator needs to be exchanged or restored to factory settings, the following
steps will adjust the settings of the pressure regulator:
Press
for a long time
F0 (adjustment of the entity)
Press
change between
and
Or take
and
to change to
Confirm with
and go back to menu
Press
F1 (adjustment of the ouput)
Press
change between
and
Press
or
until
is displayed.
Press
for change between
and
Press
or
until
is displayed.
Press
for change between [P1L] and 3.00
Press
or
until the desired number (e. g. 2.00)
Press
for change between [P1H] and 6.00
Press
or
until the desired number (e. g. 10.00)
Press
change between
and
Press
until 0.00 is displayed
Summary of Contents for ENDODRY
Page 1: ...one click SERVICE MANUAL...
Page 23: ...page 23 of 28 6 Press the Install button to install the Endodry application...
Page 26: ...page 26 of 28 6 Colour touch screen I IP 20 Display Electrical protection class IP code...
Page 28: ...page 28 of 28 7 Technical Service Information Date Inspection Signature...