Calibration screen
1) Set the measuring device on the object to be measured,
and press to execute measurement.
2/4
Step 1. Starting and ending the device and calibrating
The setup procedure for the Surface Reflectance
Analyzer is shown below.
Step 1. Starting and ending the device and calibrating
Step 2. Taking a measurement
Step 3. Installing the application software
Step 4. Starting the application software
Setup procedure
Step 2. Taking a measurement and checking the result
Quick Manual
Power button
1) Press the power button to turn on the measuring device.
2) Perform calibration immediately after the device starts.
Startup screen
3) To calibrate the measuring device, you use the
accompanying calibration unit.
Check that no dust or stains are on the standard plate,
and set the measuring device in the calibration unit.
4) Press on the measuring device to execute calibration.
When the message “Calibrating” disappears and the main
screen appears, calibration is complete.
SCAN
Measuring screen
Main screen
One-dimensional BRDF (20°)
Two-dimensional BRDF (20°)
3) Press on the main screen to display the 20° one-
dimensional BRDF.
Press again to display the 60° one-dimensional BRDF.
Press on the main screen the 20° two-dimensional BRDF
and yet again to display the 60° two-dimensional BRDF.
Gloss
Haze
Image clarity
SCAN
When the message “Measuring” disappears and the
main screen appears, measurement is complete.
Calibration unit
Standard plate
Measuring
device