– 13 –
7. Specifications
7.1 Main Specifications
Purpose
General radiography
Pixel size
160µm × 160 µm
Image matrix size
2208 × 2688 pixels
Number of pixels
Approx. 5.9 million pixels
Dynamic range
Approx. 80 dB
Resolution
3.1 lp/mm
Gray scale
12-bit, 4,096 gray scale
Environmental requirements
Operation:
Temperature: +5 to +35°C
Humidity: 30 to 75%RH (no condensation)
Storage and transportation:
Temperature: -30 to +50°C
Humidity: 10 to 60%RH (no condensation)
Atmospheric pressure: 700 to 1060hPa
Sensor Unit
Power
Supplied from the power box.
Dimensions and mass
Sensor unit
491 (W) × 477 (H) × 23 (D) mm,
4.8 kg (without sensor cable)
Sensor cable
7m type: 7 m, 0.9 kg
3m type: 3 m, 0.4 kg
Power Box
Power
100-120VAC, 50/60Hz, 1.5A
200-240VAC, 50/60Hz, 0.8A
Dimensions and mass
358 (W) × 75 (H) × 200 (D) mm,
4.2 kg