background image

Summary of Contents for Aplio-i Series

Page 1: ...CANON MEDICAL SYSTEMS CORPORATION 2017 2019 ALL RIGHTS RESERVED Disinfection Guide for Aplio i series Systems ...

Page 2: ...aintenance or preventive maintenance contracts contact your Canon Medical Systems service representative WARNING Never perform service and maintenance operations while a patient is being examined Doing so may subject the patient or operator to electric shock or cause damage to the system CAUTION 1 If preventive maintenance is performed by the user special care must be taken to ensure safety 2 If a...

Page 3: ...nits an electric shock may occur Do not use solvents such as paint thinner benzine or alcohol or abrasive cleansers to clean the system These substances may cause deterioration or discoloration of the system Cleaning the transducer Clean disinfect and sterilize the transducer referring to the operation manual provided with the transducer 2 Cleaning the transducer holder and gel holder a Wipe off a...

Page 4: ...it or apply excessive force to the LCD screen during cleaning Doing so may damage the monitor 2 Never use a stiff or hard cloth to clean the monitor Doing so may damage the monitor cover or LCD screen 3 Do not allow liquid droplets to remain on the LCD screen The droplets may cause small stains when they dry or may enter the LCD screen possibly resulting in failure a Monitor cover Clean the monito...

Page 5: ...trackball a Wipe the system using a soft cloth moistened with mild detergent b Clean around the switches or keys on the main panel using cotton buds c Wipe the transducer connector with a soft dry cloth If it is difficult to remove all the stains with a dry cloth use a soft cloth moistened with water a Turn the trackball ring in the direction of the arrow in the left figure below and remove it tog...

Page 6: ...e ring and remove any dust and dirt on the contact surface Contact surface Inner ring d Remove all foreign matter from the ball and from the interior of the case using a soft cloth or gauze Do not use sharp or hard materials to clean inside the case In addition use a microfiber cloth to clean lens cover sections left and bottom Lens cover left Lens cover bottom II I JC77 1 1 A 1 Cll l I I ...

Page 7: ...t fitted properly the ball may not rotate properly Be sure to fit the inner ring properly Tall flap Inner ring f Gently set the ring in the unit and rotate the ring so that the mark is at the position shown below Make sure that there is no foreign matter inside the trackball assembly II I JC77 1 1 A 1 Cll l I I ...

Page 8: ... ON the power and confirm that the trackball operation is normal 8 Cleaning the air filter Check and clean the air filter to prevent overheating of the system or reduction in system performance reliability due to clogging of the filter NOTICE Do not boot up the system while the air filter is removed The system may malfunction a Remove the air filter 1 location Air filter b Remove all dust from the...

Page 9: ...ves each time disinfection is performed 2 Do not disinfect the system using a method or chemical other than those specified in this manual Doing so may result in the system not being disinfected properly or in the surface of the system deteriorating or becoming discolored 3 After chemical disinfection dry the surface of the system fully 4 The efficacy of disinfectants is not guaranteed by Canon Me...

Page 10: ...Casters Footswitch Peripheral devices printer etc Refer to the documentation provided with the peripheral devices Cables Disinfection procedures Perform the disinfection work according to the instructions of a disinfection professional a Turn the system power OFF and disconnect the power cable plug from the outlet of the facility b Wear protective gloves to prevent infection Wear new sterile prote...

Reviews: