Candela Vbeam Perfecta, Platinum, Aesthetica Treatment Guidelines 

 

8502-00-0891 Revision G            Candela Proprietary                             Sheet 8 of 8 

 

 

 

If hypo or hyperpigmentation occurs treatment should not be performed until pigmentation returns 

to normal. 

Scarring (rare, possibly permanent) 

 

G. 

Determining the end of treatment

            

 

1. 

 Multiple treatments over a period of several months may be required. 

 
2. 

The provider’s judgment will determine the end of treatment. Treatment should cease when 
the desired clinical endpoint and patient result has been achieved.  

             

3. 

 Maintenance treatments may be required to maintain the desired result. 

 

H. 

Additional   Warnings and Precautions                                                                                                   

CAUTION 

A buildup of frost may occur on the distance gauge during extended treatments. Wipe the distance gauge 
with a warm moist cloth to prevent accumulation of frost 
 

CAUTION 

If crescent shaped marks, welts or wheals are noted during treatment, the fluence and or DCD settings 
should be adjusted immediately. Failure to adjust these parameters can result in subsequent crusting, 
blistering, or other adverse reactions.  Crescents may result from not holding the distance gauge properly. If 
problem persists, discontinue treatment and refer to user verification test section of the Operator’s Manual. 
 

CAUTION 

Do not use Flammable products on the skin or in the vicinity of the laser. 
 

WARNING 

When using the Vbeam laser with the Dynamic Cooling Device (DCD) in use in conjunction with an ECG 
monitoring device interference with the ECG monitoring device may result. 
 

WARNING 

Oxygen will accelerate combustion of any flammable material

.

  Oxygen should never be directed towards or 

over the laser field, extreme caution should be used. When oxygen is in use, gauze, sponges, and hair near 
the treatment area should be moistened with water.  Be aware that hairspray, gel, deodorants, and other 
grooming products may contain flammable substances. 

 

CAUTION 

Treatment to areas bearing hair may cause the hair to be singe and may cause it to ignite.  A fire 
extinguisher should be available. 
 
 

 

CRYOGEN REORDER

 - Replacement canisters of cryogen are ordered by the part number listed on the 

GentleCool

tm

 canisters shipped with laser. Use only 1000-gram size canisters for this product. 

 

Summary of Contents for Vbeam Perfecta

Page 1: ...be worn by all persons in the treatment room during laser operation Perform User Verification tests as outlined in the Operators Manual at the start of each treatment day and when the hand piece is ch...

Page 2: ...ear and become bluish pink dusky erythema May use 10 12mm spot then retreat larger vessels immediately with smaller spot size Press the lens firmly onto the skin until the skin turns pale Clinical end...

Page 3: ...fluences and DCD for neonates and infants Clinical endpoint purpura Use lower fluences on neck and extremities Ectatic vessels associated with hypertrophic PWS may be double pulsed Warts Non Facial F...

Page 4: ...DCD spray should be sufficient to cover the treatment area and provide epidermal cooling in opposition to the heat generated from the laser It is important to maintain both skin protection and patien...

Page 5: ...0920 23 0814 e Seizure disorders Do not treat patients with a history of light triggered seizures f Medications Daily anticoagulation therapy iron supplements herbal supplements such as ginkgo ginseng...

Page 6: ...cted treatment response is noted B Procedure for establishing the Purpura Free Threshold 1 Select spot size pulse duration fluence and DCD settings 2 Apply escalating fluences until purpura appears th...

Page 7: ...e used to draw a treatment area grid When treating anal area place wet gauze over the anus first Methane gas is flammable 8 The Cleaning and Disinfecting Policy is outlined in Section 6 of the Operato...

Page 8: ...in subsequent crusting blistering or other adverse reactions Crescents may result from not holding the distance gauge properly If problem persists discontinue treatment and refer to user verification...

Reviews: