Candela Corporation
M
GentleLASE PLUS
Installation Procedure
8503-01-0821, Revision A
Candela Corporation Proprietary
Page 5 of 8
Initial Cryogen Canister Installation
Figure 1 Canister and Delivery System Locations
1.
Remove the DCD cryogen canister from the packaging box. Be sure to read all
labels on the canister before installing into the system.
2.
Insert canister threaded side down into the opening on the top left side of the laser
system. Turn canister clockwise until hand tight. A small amount of cryogen
might vent out. The sound of the escaping gas may be quite loud. Only after the
canister is fully seated will the gas stop venting out.
3.
Place the cylindrical canister cap over the cryogen canister.
Display
Panel
Laser Stop
Switch
Ready
Lamp
Calibration
Port
Valve
Control
Receptacle
Cryogen
Line
Receptacle
Fiberoptic
Receptacle