TB4 and TB4MM Tipping Bucket Rain Gages 

Resolution:

  

0.254 mm (0.01 in) TB4 

 

 

0.008 in (0.2 mm) TB4MM 

Environmental Conditions 
 

Temperature:

 

0 to 70 °C 

 

Humidity:

 

0 to 100% 

Operating Temperature:

 

–20 to 70 °C 

Siphon Capacity:

 

0.3 mm (0.012 in) 

Contact:

 

Dual reed switch 

Capacity:

 

12 VA (0.5 A maximum) 

Base Material:

 

Thermoplastic 

Size 
 

Weight:

 

2 kg (4.41 lb)  

 

Height:

 

330 mm (13 in) 

 

Diameter:

 

200 mm (7.9 in) 

7.  Installation 

If you are programming your datalogger with 

Short Cut

, skip Section 

7.1

Wiring

 

(p. 5)

, and Sectio

7.2

Datalogger Programming

 

(p. 6)

Short Cut 

does 

this work for you. See Section 

4

QuickStart

 

(p. 2)

, for a 

Short Cut 

tutorial. 

The 260-953 Alter-Type Wind Screen’s siting information and 
installation procedure are provided in our 

260-953 manual

. 

7.1  Wiring 

Connections to Campbell Scientific dataloggers are given in TABLE 

7-1

 

TABLE 7-1.  Wire Color, Wire Function, Datalogger Connection  

 

Wire 

Color 

 

Wire 

Function 

Datalogger Connection 

Terminal for Pulse 

Channel Input 

Datalogger Connection 

Terminal for Control 

Port Input

1

 

Black 

Rain signal 

P

P_SW

, or 

U

1

 

(pulse channel)  

(control port)  

White 

Rain signal 

reference 

AG 

or 

 

(analog ground) 

5 V 

(on datalogger)  

Clear 

Shield 

AG 

or 

 

(analog ground) 

AG

 or 

 

(analog ground) 

1

U

 channels are automatically configured by the measurement instruction.

 

 
A long cable has an appreciable capacitance between the lines. A built-up 
charge could cause arcing when the switch closes, shortening switch life. A 

NOTE 

Summary of Contents for TB4

Page 1: ...INSTRUCTION MANUAL TB4 and TB4MM Rain Gages Revision 4 16 C o p y r i g h t 1 9 9 5 2 0 1 6 C a m p b e l l S c i e n t i f i c I n c...

Page 2: ......

Page 3: ...ucts to CSI CSI will return such Products by surface carrier prepaid within the continental United States of America To all other locations CSI will return such Products best way CIP port of entry per...

Page 4: ...g container Campbell Scientific s shipping address is CAMPBELL SCIENTIFIC INC RMA _____ 815 West 1800 North Logan Utah 84321 1784 For all returns the customer must fill out a Statement of Product Clea...

Page 5: ...beginning work Wear a hardhat and eye protection and take other appropriate safety precautions while working on or around tripods and towers Do not climb tripods or towers at any time and prohibit cl...

Page 6: ......

Page 7: ...tion 6 7 3 Siting 7 7 4 Mounting 7 7 5 Leveling 8 8 Troubleshooting and Maintenance 9 8 1 Troubleshooting 9 8 2 Maintenance 10 8 2 1 Dismantling for Cleaning 10 8 2 2 Reassembling the TB4 13 8 3 Calib...

Page 8: ...of the TB4 9 8 1 Components of TB4 Base 10 8 2 Dismantling the Filter Siphon Assembly 11 8 3 Filter Siphon Assembly 12 8 4 Reassembling the TB4 13 Table 7 1 Wire Color Wire Function Datalogger Connec...

Page 9: ...e precision instruments Please handle them with care The black outer jacket of the cable is Santoprene rubber This compound was chosen for its resistance to temperature extremes moisture and UV degrad...

Page 10: ...datalogger wiring terminals Short Cut is available as a download on www campbellsci com and the ResourceDVD It is included in installations of LoggerNet PC200W PC400 or RTDAQ Use the following proced...

Page 11: ...Rain box and selecting one of the other options The default rainfall per tip value of 0 01 inches should be used if the sensor is a TB4 purchased at Campbell Scientific U S office Select 0 2 mm if the...

Page 12: ...tips and activates a reed switch The switch closure is recorded by the datalogger pulse channel When the bucket tips the water drains out the screened fittings in the base of the gage The rain gages...

Page 13: ...ial The 260 953 Alter Type Wind Screen s siting information and installation procedure are provided in our 260 953 manual 7 1 Wiring Connections to Campbell Scientific dataloggers are given in TABLE 7...

Page 14: ...fter they are imported and edited in CRBasic Editor A Short Cut tutorial is available in Section 4 QuickStart p 2 If you wish to import Short Cut code into CRBasic Editor to create or add to a customi...

Page 15: ...surface Each pad includes a hole that fits a 3 8 inch M8 bolt The three holes form a 234 mm 9 21 in diameter bolt circle Campbell Scientific offers the CM240 mounting bracket for installing and leveli...

Page 16: ...emove the housing assembly from the base by loosening the three housing screws and lifting the housing upward 2 Adjust the three leveling screws on the CM240 bracket to level the gage FIGURE 7 4 A bul...

Page 17: ...tion of Hazardous Material and Decontamination form Refer to the Assistance page at the beginning of this manual for more information 8 1 Troubleshooting Symptom No Precipitation 1 Check that the sens...

Page 18: ...ach bucket tip 8 2 1 Dismantling for Cleaning The following items should be checked regularly for cleanliness Filter FIGURE 8 3 Siphon FIGURE 8 3 Tipping bucket assembly FIGURE 8 1 Top surface of adju...

Page 19: ...pulling the siphon see FIGURE 8 2 Do not twist the filter siphon assembly while pushing and pulling FIGURE 8 2 Dismantling the Filter Siphon Assembly CAUTION Do not twist while pushing and pulling Pu...

Page 20: ...y doing the following FIGURE 8 3 a Unscrew nut b Lightly press stem down on surface until stem pops out of siphon body c Remove stem from siphon body d Unscrew cap e Clean all items FIGURE 8 3 Filter...

Page 21: ...assembly while putting it back into place FIGURE 8 4 Reassembling the TB4 5 Place the housing assembly back onto the base and tighten the three screws that secure the housing onto the base 8 3 Calibr...

Page 22: ...314 cc 19 16 in3 of water over a 15 minute period into the collection funnel This volume of water is equal to 39 inches of rainfall 10 mm 4 After the water has passed through the rain gage the tipping...

Page 23: ...ram and exit Short Cut Make note of the file name used when saving the Short Cut program 2 Open CRBasic Editor 3 Click File Open Assuming the default paths were used when Short Cut was installed navig...

Page 24: ......

Page 25: ...ain Gage measurement Rain_mm PulseCount Rain_mm 1 1 2 0 0 254 0 Call Data Tables and Store Data CallTable TB4_mm NextScan EndProg B 1 2 CR200 X Series Pulse Channel Example CRBasic Example B 2 CR200 X...

Page 26: ...0 Min 0 Totalize 1 Rain_mm FP2 0 EndTable Main Program BeginProg Scan 1 Sec 1 0 PulseCount Rain_mm 1 18 2 0 254 0 Black wire connected to C8 CallTable Rain NextScan EndProg B 2 2 CR200 X Series Contro...

Page 27: ......

Page 28: ...ua Apinag s nbr 2018 Perdizes CEP 01258 00 S o Paulo SP BRASIL www campbellsci com br vendas campbellsci com br Campbell Scientific Canada Corp 14532 131 Avenue NW Edmonton AB T5L 4X4 CANADA www campb...

Reviews:

Related manuals for TB4