SC-IRDA Infra-red Communications Interface

6

4.2.1  PC200W Version 1.2 and PC208W Version 3 (or later)

Both of these programs will work with the SC-IRDA on all but the slowest
PCs. PC200W only supports COM 1 to COM 4, and so you must ensure the
virtual IRDA serial port maps to one of these numbered ports. PC208W does
not have this limitation.

Both of these programs will generate a warning message in the low level log
files when the serial port is first opened. This can be ignored.

PC200W version 2.0 does not work with the SC-IRDA. Please
contact Campbell Scientific to check if a later version is
available which includes support.

4.2.2  LoggerNet

Version 2.1 of LoggerNet works with the SC-IRDA but requires the addition of
a Generic Modem to be inserted between the virtual COM port and the
datalogger in the network map. The only non-default setting required in the
Generic modem setup is the dial string which should be entered as "D5000".
This setting forces LoggerNet to wait for 5 seconds after opening the serial port
before it tries to communicate. Later versions of LoggerNet should not need
this setting.

Users of Windows 98 or ME may encounter a problem with LoggerNet not
allowing use of the virtual COM port assigned to the IRDA interface. This is
due to inconsistencies in the way valid COM ports are reported to applications.
Please contact Campbell Scientific for further advice on how to overcome this
problem.

The version of SMS supplied with LoggerNet 2.1 will not allow
communication to a storage module connected in parallel to a datalogger.
Version 4.6 of SMS, which will be shipped with LoggerNet 2.2, is needed to
allow for the extra start up delay associated with the IRDA interface. Please
contact Campbell Scientific if you need a copy prior to the formal release of
this version.

4.2.3  PC9000

All versions of PC9000 should work with the SC-IRDA interface by setting the
program to operate via the virtual COM port assigned to the SC-IRDA. As with
all PC programs, the serial port speed set on the PC does not actually control
the real baud rate used to communicate with the datalogger. With PC9000 this
is fixed at 9600 baud and cannot be changed. This baud rate may be a severe
limitation if you need to transfer large volumes of data.

4.3  Optimizing the Datalogger Baud Rate

The speed of connection between the datalogger and SC-IRDA interface is set
at 9600 baud by default, independently of any baud rate settings you enter in
the PC software. This speed ensures "out of the box" operation with all
Campbell Scientific dataloggers.

The interface itself has the capability to run at different baud rates, e.g. 1200,
9600 or 19200 baud. Using the higher speed setting will increase throughput of

NOTE

Summary of Contents for SC-IRDA

Page 1: ...INSTRUCTION MANUAL SC IRDA Infrared Communications Interface Revision 6 04 C o p y r i g h t c 2 0 0 3 2 0 0 4 C a m p b e l l S c i e n t i f i c I n c ...

Page 2: ... of all other warranties expressed or implied including warranties of merchantability or fitness for a particular purpose CAMPBELL SCIENTIFIC INC is not liable for special indirect incidental or consequential damages Products may not be returned without prior authorization The following contact information is for US and International customers residing in countries served by Campbell Scientific In...

Page 3: ...l 1 2 4 Datalogger Communication Modes 2 2 5 Supported Software Packages 2 3 Installation 2 3 1 Hardware 2 3 2 Using PC Software Drivers 4 4 Operating Instructions 5 4 1 Palm OS Handheld Software 5 4 2 PC Software 5 4 2 1 PC200W Version 1 2 and PC208W Version 3 or later 6 4 2 2 LoggerNet 6 4 2 3 PC9000 6 4 3 Optimizing the Datalogger Baud Rate 6 4 4 CSOS Datalogger Operating System Download Utilit...

Page 4: ...This is a blank page ...

Page 5: ...icating with some dataloggers the CR510 CR10X and CR23X with new standard operating systems that allows it to be connected at the same time as other modem devices for example a telephone modem thus avoiding the need for any extra port switching device 2 Specifications 2 1 Electrical Standby Current 0 17mA typical Active communicating 6 2mA typical current drawn from the 5V supply provided by the d...

Page 6: ...ggerNet Version 2 1a PConnect for Palm PCs version 2 1 or later PC200W version 1 2 but not version 2 0 PC208W Version 3 0 or later PC9000 at 9600 baud only Storage Module software Version 4 6 and later PC Operating Systems Support Windows 98 and ME directly Windows 2000 and XP require a third party virtual serial port driver supplied Please check with Campbell Scientific for availability of compat...

Page 7: ...st or rain on the sensor can normally be gently wiped off with a soft cloth before communication is started Exposure of the sensor to bright sunlight may reduce the maximum spacing between computer and the interface Exposure to strong artificial lights which have a high frequency electronic ballast may also reduce the working range In addition they may cause the SC IRDA to stay awake and thus draw...

Page 8: ...abled for any reason in the PC s BIOS Please refer to your system documentation and Microsoft s installation help system for further information To allow you to use the virtual serial port you need to know which COM port number has been assigned to the IRDA port This information is normally found using the Windows Control Panel either via General System Information an IRDA setup link or via a task...

Page 9: ...tocols will ensure no loss of data However if the connection is broken for more than a few seconds the SC IRDA may shut down This may then require PConnect to be terminated restarted and the connection re initialized before further data can be collected or displayed Some of the older slower Palm PCs are not able to keep up with the speed of transfer of data when using the option to read a datalogg...

Page 10: ... due to inconsistencies in the way valid COM ports are reported to applications Please contact Campbell Scientific for further advice on how to overcome this problem The version of SMS supplied with LoggerNet 2 1 will not allow communication to a storage module connected in parallel to a datalogger Version 4 6 of SMS which will be shipped with LoggerNet 2 2 is needed to allow for the extra start u...

Page 11: ...be used if another SDC device is used at the same time To send the string PC208W and LoggerNet require you to add a generic modem between the COM port and the datalogger in the network map A command is then entered in the dial string for the generic modem as shown below D5000T SETUP 2 D5000 instructs the program to wait 5000 milliseconds after opening the serial port before transmitting the string...

Page 12: ...SC IRDA Infra red Communications Interface 8 This is a blank page ...

Page 13: ...This is a blank page ...

Page 14: ... campbellsci com br suporte campbellsci com br Campbell Scientific Canada Corp CSC 11564 149th Street NW Edmonton Alberta T5M 1W7 CANADA www campbellsci ca dataloggers campbellsci ca Campbell Scientific Ltd CSL Campbell Park 80 Hathern Road Shepshed Loughborough LE12 9GX UNITED KINGDOM www campbellsci co uk sales campbellsci co uk Campbell Scientific Ltd France Miniparc du Verger Bat H 1 rue de Te...

Reviews: