PWC100 Present Weather Sensor Calibrator
Step 4: Position 2: Left
•
Move the calibrator into the left hand position by unscrewing the calibrator
mounting screws slightly and sliding the calibrator fully to the left. Then
retighten the screws once in position.
•
Visually check that the calibrator is located correctly and securely.
•
LED
L
should now be flashing slowly and all other LEDs should be off.
•
Once you are ready, press
START
.
•
The test is now automated and the disk should start rotating. LED
L
will
stay illuminated until this stage of the test is complete.
•
Wait for the calibration disk to stop rotating. This test will take
approximately two and a half minutes.
•
Upon successful completion of the test, LED
R
will be flashing slowly and
LED
L
will be turned off.
7