Step 3: Position 1: Center position
Attaching the velocity calibration unit
•
Fix the PWC100 velocity calibrator unit to the center position on the PWS
using two of the countersunk fixing screws provided
•
Check that the velocity calibrator is located correctly and securely.
•
LED
C
should now be flashing slowly and all other LEDs should be off.
•
Once you are ready, press
START
.
•
The test is now automated and the disk should start rotating. LED
C
will
stay illuminated until this stage of the test is complete.
•
Wait for the calibration disk to stop rotating. This test will take
approximately two and a half minutes.
•
Upon successful completion of the test, LED
L
will be flashing slowly and
LED
C
will be turned off.