Appendix C. EasyFlux
®
DL CR6KH20
C-7
TABLE C-6. Default Wiring for Radiation Measurements Option 2
Sensor
Quantity
Wire Description
Color
Terminal
Current Excite
Red (cbl 2)
VOLT116 X1
Current Return
Blue (cbl 2)
VOLT116
⏚
Shields
Clear
VOLT116
⏚
CNR4 4-Way
Radiometer
0 or 1
Pyranometer Up Signal
Red
VOLT116 Diff 9H
Pyranometer Up Reference
Blue
1/
VOLT116 Diff
9L
1/
Pyranometer Down Signal
White
VOLT116 Diff
10H
Pyranometer Down
Reference
Black
1/
VOLT116 Diff
10L
1/
Pyrgeometer Up Signal
Grey
VOLT116 Diff
11H
Pyrgeometer Up Reference
Yellow
1/
VOLT116 Diff
11L
1/
Pyrgeometer Down Signal
Brown
VOLT116 Diff
12H
Pyrgeometer Down
Reference
Green
1/
VOLT116 Diff
12L
1/
Thermistor Signal
White
VOLT116 Diff 4H
Thermistor V Excite
Red
VOLT116 X1
Thermistor Reference
Black
VOLT116
⏚
Shields
Clear
VOLT116
⏚
1/
Jumper to
⏚
with user-supplied wire
A CNF4 Ventilation and Heater Unit may be used with the CNR4 4-way
Radiometer for more accurate radiation measurements. The CNF4 requires a
solid state relay to control the ventilator and heater. The A21REL-12
4-Channel Relay Driver, sold separately, is recommended. Install the A21REL-
12 inside the system enclosure near the VOLT116 and data logger. TABLE
lists the wiring connections needed to power and control the A21REL-12;
a CABLE3CBL-1 or similar 3-conductor 22 AWG cable is recommended for
connections from the A21REL-12 to the VOLT116, and a CABLEPCBL-1 or
similar 16 AWG 2-conductor power cable is recommended for power
connections from the A21REL-12 to system 12V power source. TABLE
lists the wiring for the CNF4.