background image

Appendix A.  Calibrating KH20 

A.1  Basic Measurement Theory 

The KH20 uses an empirical relationship between the absorption of the light 
and the material through which the light travels.  This relationship is known as 
the Beer’s law, the Beer-Lambert law, or the Lambert-Beer law.  According to 
the Beer’s law, the log of the transmissivity is anti-proportional to the product 
of the absorption coefficient of the material, 

k

, the distance the light travels, 

x

and the density of the absorbing material, 

ρ

.  The KH20 sensor uses the UV 

light emitted by the krypton lamp: major line at 123.58 nm and the minor line 
at 116.49.  As the light travels through the air, both the major line and the 
minor line are absorbed by the water vapor present in the light path.  This 
relationship can be rewritten as follows, where 

k

w

 is the absorption coefficient 

for water vapor, 

x

 is the path length for the KH20 sensor, and 

ρ

w

 is the water 

vapor density. 

w

w

x

k

e

T

ρ

=

 A-1 

If we express the transmissivity, 

T

, in terms of the light intensity before and 

after passing through the material as measured by the KH20 sensor, 

V

 and 

V

0

respectively, we obtain the following equation. 

w

w

x

k

e

V

V

ρ

=

0

 A-2 

Taking the natural log of both sides, and solving for the density, 

ρ

w

, yields the 

following equation. 

)

ln

(ln

1

0

V

V

x

k

w

w

=

ρ

 A-3 

If the path length, 

x

, and the absorption coefficient for water, 

k

w

 are known, it 

becomes possible to measure the water vapor density 

ρ

w

, by measuring the 

signal output, 

V

, from KH20. 

A.2  Calibration of KH20 

The KH20 calibration process is to find the absorption coefficient of water 
vapor, 

k

w

.  To do this, we rewrite the equation A-3, and solve for ln(V).   

0

ln

ln

V

x

k

V

w

w

+

=

ρ

 A-4 

It now becomes obvious from the equation A-4 that there is a linear 
relationship between the natural log of the KH20 measurement output, ln

V

, and 

the water vapor density, 

ρ

w

.  Figure A-1 shows the plot of the equation A-4 

after we ran a KH20 over a full calibration vapor range. 

A-1 

Summary of Contents for KH20

Page 1: ...KH20 Krypton Hygrometer 2 10 C o p y r i g h t 2 0 1 0 C a m p b e l l S c i e n t i f i c I n c...

Page 2: ...customers residing in countries served by Campbell Scientific Inc directly Affiliate companies handle repairs for customers within their territories Please visit www campbellsci com to determine whic...

Page 3: ...e Electronics Box 3 4 Wiring 5 5 Datalogger Programming 6 5 1 KH20 Calibration 6 5 2 Example 1 CR3000 Program to Measure Water Vapor Fluctuations 7 6 Maintenance and Calibration 9 6 1 Visual Inspectio...

Page 4: ...f Contents ii Tables 1 Datalogger Connections for Differential Measurement 5 2 KH20 Calibration Ranges 6 A 1 Linear Regression Results for KH20 ln mV vs Vapor Density A 2 A 2 Final Calibration Values...

Page 5: ...inside a rain shield CSI p n 14113 that protects them from moisture The KH20 is suitable for long term continuous outdoor applications The KH20 sensor is comprised of two main parts the sensor head an...

Page 6: ...13 Horizontal Mounting Boom 20 inch 3 4 IPS threaded aluminium pipe CSI p n 3874 Nu rail 3 4 x 3 4 CSI p n 1017 Allen Wrench 5 32 CSI p n 4697 3 Installation 3 1 Siting When installing the KH20 sensor...

Page 7: ...the two tubes is positioned on top as shown in Figure 1 Gather any loose cables and tie them up using cable ties onto the tripod or tower mast FIGURE 1 Mounting KH20 onto a tripod 3 2 3 Mounting the...

Page 8: ...s box later on FIGURE 2 Proper mounting position of the electronics box If the electronics box is not pushed all the way up during mounting you will not have enough room to attach the connectors to th...

Page 9: ...n the two pan head screws to secure it in place 6 Gather any loose cables and tie them up using cable ties onto the tripod or tower mast 4 Wiring The KH20 sensor is shipped with a power signal cable w...

Page 10: ...cients are calculated See Appendix A for more information on KH20 calibration TABLE 2 KH20 Calibration Ranges Ranges Vapor Density g m3 Full Vapor Range 2 19 Dry Vapor Range 2 9 5 Wet Vapor Range 8 25...

Page 11: ...f these constants and enter the appropriate values found from the calibration sheet of the KH20 Unit Definitions Units Units ln_mV ln mV natural log of the KH20 millivolts mV millivolts rho_w g m 3 Wi...

Page 12: ...AL Min 10 Minimum 1 batt_volt FP2 False False Average 1 panel_temp FP2 False Average 2 kh 1 IEEE4 False EndTable Raw time series data DataTable ts_data True 1 DataInterval 0 SCAN_INTERVAL mSec 100 Sam...

Page 13: ...the test explained below First make sure the UV light is emitted from the source tube To do this you may look into the source tube the longer of the two tubes and you should see a bright blue light e...

Page 14: ...o see if the source tube window has been scaled get a clean dry cotton swab and slide it across the source tube window The scale is not visible to the naked eye but if the window is scaled you will fe...

Page 15: ...ngth for the KH20 sensor and w is the water vapor density w wx k e T A 1 If we express the transmissivity T in terms of the light intensity before and after passing through the material as measured by...

Page 16: ...mV vs Vapor Density Description Values Slope xkw 0 205 Y Intercept ln V0 8 033 If we substitute these values along with the measured lnV into equation A 3 we can obtain the water vapor density w Campb...

Page 17: ...he data shown in Table A 2 is from an actual KH20 TABLE A 2 Final Calibration Values for KH20 Vapor Range g m3 Slope xkw Y Intercept ln V0 Coefficient kw Full Range 1 74 19 25 0 205 3087 0 144 Dry Ran...

Page 18: ...Appendix A Calibrating KH20 A 4...

Page 19: ......

Page 20: ...a Corp CSC 11564 149th Street NW Edmonton Alberta T5M 1W7 CANADA www campbellsci ca dataloggers campbellsci ca Campbell Scientific Centro Caribe S A CSCC 300 N Cementerio Edificio Breller Santo Doming...

Reviews: