KH20 Krypton Hygrometer 

should see an immediate decrease in the voltage reading, and it should go close 
to zero.  No noticeable change in the voltage output, when the optical path is 
completely blocked, indicates a problem in the detector tube.  If the decrease in 
the voltage reading takes place but the reading remains below 50 mV, when 
the paper is removed from the optical path, the source tube may be at fault.  
Obtain an RMA from Campbell Scientific, Inc. and send the unit in for repair. 

6.4  Managing the Scaling of KH20 

The KH20 cannot be used to measure an absolute concentration of water vapor, 
because of scaling on the source tube windows caused by disassociation of 
atmospheric continuants by the ultra violet photons (Campbell and Tanner, 
1985 and Buck, 1976).  The rate of scaling is a function of the atmospheric 
humidity.  In a high humidity environment, scaling can occur within a few 
hours.  That scaling attenuates the signal and can cause shifts in the calibration 
curve.  However, the scaling over a typical flux averaging period is small.  
Thus, water vapor fluctuation measurements can still be made with the 
hygrometer.   

To see if the source tube window has been scaled, get a clean, dry cotton swab 
and slide it across the source tube window.  The scale is not visible to the 
naked eye, but if the window is scaled, you will feel a slight but noticeable 
resistance while you slide the swab across the window.  There will be little 
resistance if the window is not scaled.  If you determine the window is scaled, 
you can clean it with a wet cotton swab.   

Use distilled water and a clean cotton swab to clean the scaled window.  After 
cleaning the window, slide a clean, dry swab across the window to confirm the 
scale has been removed. 

You can use the water vapor absorption coefficient for scaled 
window from the calibration sheet if the window will be allowed 
to scale during measurements. 

NOTE 

6.5  Older KH20 

The older version of the KH20 used to suffer permanent damage when exposed 
to water, due to corrosion and loss of vacuum within the tubes.  The new 
version of the KH20 has employed a new sealing around the window to 
prevent this type of damage. 

6.6  Calibration 

For quality assurance of the measured data, Campbell Scientific, Inc. 
recommends the KH20 be recalibrated every two years.  Please contact 
Campbell Scientific, Inc. to obtain an RMA number for recalibration.   

For more information on the calibration process, refer to Appendix A of this 
manual.  

10 

Summary of Contents for KH20

Page 1: ...KH20 Krypton Hygrometer 2 10 C o p y r i g h t 2 0 1 0 C a m p b e l l S c i e n t i f i c I n c...

Page 2: ...customers residing in countries served by Campbell Scientific Inc directly Affiliate companies handle repairs for customers within their territories Please visit www campbellsci com to determine whic...

Page 3: ...e Electronics Box 3 4 Wiring 5 5 Datalogger Programming 6 5 1 KH20 Calibration 6 5 2 Example 1 CR3000 Program to Measure Water Vapor Fluctuations 7 6 Maintenance and Calibration 9 6 1 Visual Inspectio...

Page 4: ...f Contents ii Tables 1 Datalogger Connections for Differential Measurement 5 2 KH20 Calibration Ranges 6 A 1 Linear Regression Results for KH20 ln mV vs Vapor Density A 2 A 2 Final Calibration Values...

Page 5: ...inside a rain shield CSI p n 14113 that protects them from moisture The KH20 is suitable for long term continuous outdoor applications The KH20 sensor is comprised of two main parts the sensor head an...

Page 6: ...13 Horizontal Mounting Boom 20 inch 3 4 IPS threaded aluminium pipe CSI p n 3874 Nu rail 3 4 x 3 4 CSI p n 1017 Allen Wrench 5 32 CSI p n 4697 3 Installation 3 1 Siting When installing the KH20 sensor...

Page 7: ...the two tubes is positioned on top as shown in Figure 1 Gather any loose cables and tie them up using cable ties onto the tripod or tower mast FIGURE 1 Mounting KH20 onto a tripod 3 2 3 Mounting the...

Page 8: ...s box later on FIGURE 2 Proper mounting position of the electronics box If the electronics box is not pushed all the way up during mounting you will not have enough room to attach the connectors to th...

Page 9: ...n the two pan head screws to secure it in place 6 Gather any loose cables and tie them up using cable ties onto the tripod or tower mast 4 Wiring The KH20 sensor is shipped with a power signal cable w...

Page 10: ...cients are calculated See Appendix A for more information on KH20 calibration TABLE 2 KH20 Calibration Ranges Ranges Vapor Density g m3 Full Vapor Range 2 19 Dry Vapor Range 2 9 5 Wet Vapor Range 8 25...

Page 11: ...f these constants and enter the appropriate values found from the calibration sheet of the KH20 Unit Definitions Units Units ln_mV ln mV natural log of the KH20 millivolts mV millivolts rho_w g m 3 Wi...

Page 12: ...AL Min 10 Minimum 1 batt_volt FP2 False False Average 1 panel_temp FP2 False Average 2 kh 1 IEEE4 False EndTable Raw time series data DataTable ts_data True 1 DataInterval 0 SCAN_INTERVAL mSec 100 Sam...

Page 13: ...the test explained below First make sure the UV light is emitted from the source tube To do this you may look into the source tube the longer of the two tubes and you should see a bright blue light e...

Page 14: ...o see if the source tube window has been scaled get a clean dry cotton swab and slide it across the source tube window The scale is not visible to the naked eye but if the window is scaled you will fe...

Page 15: ...ngth for the KH20 sensor and w is the water vapor density w wx k e T A 1 If we express the transmissivity T in terms of the light intensity before and after passing through the material as measured by...

Page 16: ...mV vs Vapor Density Description Values Slope xkw 0 205 Y Intercept ln V0 8 033 If we substitute these values along with the measured lnV into equation A 3 we can obtain the water vapor density w Campb...

Page 17: ...he data shown in Table A 2 is from an actual KH20 TABLE A 2 Final Calibration Values for KH20 Vapor Range g m3 Slope xkw Y Intercept ln V0 Coefficient kw Full Range 1 74 19 25 0 205 3087 0 144 Dry Ran...

Page 18: ...Appendix A Calibrating KH20 A 4...

Page 19: ......

Page 20: ...a Corp CSC 11564 149th Street NW Edmonton Alberta T5M 1W7 CANADA www campbellsci ca dataloggers campbellsci ca Campbell Scientific Centro Caribe S A CSCC 300 N Cementerio Edificio Breller Santo Doming...

Reviews: