Appendix B. Serial Communications with the CSAT3
B-5
TABLE B-6. Converting Speed of Sound from Word 3
Conversion
Output Resolution
(mm s
-1
)
(
°
C @ 25
°
C)
speed of sound (m s
-1
) = word 3
∗
0.001 + 340.0
1 0.0017
B.4 Word 4
Word 4 contains diagnostic information and the reporting range of u
x
, u
y
, and
u
z
wind speed. The diagnostic flags, bits b15 through b12, describe the status
of the anemometer (TABLE B-7). A flag is on if the bit is set high. All the
flags are on while the anemometer is acquiring signals or is in a lost trigger
state (TABLE B-8).
TABLE B-7. Decoding the Diagnostic Flags from Word 4
Bit Set High
Description
Comment
b15
Difference in the speed of sound between
the three non orthogonal axes is greater
than 2.360 m s
-1
(~ 4°C @ 25°C)
The anemometer path length may have been altered
or the anemometer head is broken.
b14
Poor signal lock
There may be an obstruction in the anemometer path
or on the transducer face.
b13
Sonic signal amplitude too high
An obstruction, in the anemometer path or on the
transducer face, was removed.
b12
Sonic signal amplitude too low
There may be an obstruction in the anemometer path
or on the transducer face.
Bits b11 through b06 are used in determining the range of the wind speed
reported in words 0 through 2.
Bits b05 through b00 are a diagnostic counter. This counter increments once
per measurement. The minimum value is 0 and the maximum is 63. The
counter wraps from 63 to 0 when the maximum value is incremented by one.
TABLE B-8. Special Case – Not a Number
Not a Number
Word 0
Word 1
Word 2
Word 3
Word 4
Conditions
hex
base 10
hex
base 10
hex
base 10
hex
base 10
hex
base 10
Not
Available 8000 -32768 8000 -32768 8000 -32768 8000 -32768 F03F 61503
Lost
Trigger 8000 -32768 8000 -32768 8000 -32768 8000 -32768 F000 61440
B.5 CSAT3 RS-232 Status Format
The CSAT3 returns a 10 ASCII byte status in response to a “S” or “P”
command. This status provides information on the CSAT3’s Trigger Source
and the Execution Parameter. It also indicates if the analog outputs are on/off,
if the CSAT3 is reporting good data, and serial number of the CSAT3. This
software serial number matches those on the anemometer head and electronics
box.
Summary of Contents for CSAT3
Page 2: ......
Page 56: ...Appendix A CSAT3 Orientation A 4 ...
Page 66: ...Appendix C CSAT3 Measurement Theory C 4 ...
Page 70: ...Appendix D SDM Communications and Long Signal Cables D 4 ...
Page 71: ......