CS230 Temperature Profiler
6
temperatures can be used to monitor specific seasons or periods of measure,
without having to review the entire data set. The lifetime minimum and
maximum temperatures are used for maintenance and warranty records.
7.1 Wiring
Connections for the SGB3 and CS230 to Campbell Scientific dataloggers are
. The SGB3 is required to protect against
electrical surges. The SGB3 connects to the datalogger using the 2-ft cable
shipped with the CS230.
FIGURE 7-1. SGB3 3-Line Surge Protector
TABLE 7-1. CS230 Connection to SGB3
Color
Description
CS230
Red
Power
L1
Green
SDI-12 Signal
L2
Not Used
L3
Black
Power Ground
G
Clear
Shield
G
TABLE 7-2. SGB3 Connection to Campbell Scientific Dataloggers
Color SGB3 Description
Datalogger Connection Terminal
Red
L1
12V
Green
L2
Control Port
1
or U configured for SDI-12
2
Black
G
G
Clear
G
⏚
1
Dedicated SDI-12 port of CR5000
2
U channels are automatically configured by the measurement instruction.