![Campbell CS230 Field Maintenance And Instruction Manual Download Page 11](http://html1.mh-extra.com/html/campbell/cs230/cs230_field-maintenance-and-instruction-manual_3745991011.webp)
CS230 Temperature Profiler
2
The CS230 is suited for a wide variety of applications and environments. The
completely sealed probe assembly and external probes permits the CS230 to be
used in roadbeds, soils, and water (snow and ice).
Examples of some applications include spring load adjustment, frost and
permafrost monitoring, soil and water temperature profiling, and snowpack
temperature profiling.
The purpose of the SGB3 is to provide adequate surge protection for the CS230
Temperature Profiler. The case of the SGB3 is suited for mounting to a back
plate with 1 inch on center spacing.
5. Specifications
Features:
•
Accurate and stable measurements
•
Each sensor is individually addressed and referenced to its depth
•
Low power consumption
•
Digital SDI-12 output
•
Compatible with the following dataloggers: CR200(X) Series, CR800
Series, CR1000, CR3000, CR5000, CR510, CR10(X), CR23X
5.1 SGB3 3-Line Surge Protector
Operating Range:
–55 to 85
°
C
Maximum Voltage:
±28 Vdc / 20Vac
(L1, L2, L3 with respect to G terminals)
Maximum Current:
2 A per terminal, 4 A total
(requires both ground terminals for return
current)
Maximum Rated Surge:
1200 Amps (8/20 us)
5.2 CS230 Temperature Profiler
Operating Range:
–55 to 85
°
C
Accuracy
Typical:
±0.
2
°
C over –40 to 85
°
C,
Worst Case:
±0.4 °
C over –40 to 85
°
C;
±0.5 °
C over –55 to –40
°
C
(includes lifetime drift of sensor)
Resolution:
0.0078
°
C
Measurement Update Interval:
1 s (automatic), occurs in quiescent mode
Warm-up Time:
10 seconds
Summary of Contents for CS230
Page 2: ......
Page 3: ......
Page 7: ......
Page 29: ...CS230 Temperature Profiler 20...
Page 30: ......