Overview
OV-6
Normally thermocouple measurements would be made on the CR9050 Analog
Input Module (
±
5 Volt) because of its greater resolution, however they can be
made on the CR9055 if the
±
50 V common mode range is necessary.
Thermocouple, Differential Voltage (TCDiff)
Thermocouple, Single-Ended Voltage (TCDiff)
9055 50V ANALOG INPUT
SE
DIF
1
1
H
2
L
2
3
H
4
L
3
5
H
6
L
4
7
H
8
L
5
9
H
10
L
6
11
H
12
L
7
13
H
14
L
8
15
H
16
L
9
17
H
18
L
10
19
H
20
L
11
21
H
22
L
12
23
H
24
L
13
25
H
26
L
14
27
H
28
L
MADE IN USA
FIGURE OV1-10. CR9055
CR9060 Excitation Module
The CR9060 Excitation Module has six continuous analog outputs with
individual digital-to-analog converters for PID Algorithm, waveform
generation, and excitation for bridge measurements. Ten switched excitation
channels provide precision voltages for bridge measurements. Each analog
output will provide up to 50 mA between
±
5 V. Also includes eight digital
control outputs (0 V low, 5 V high).
MEASUREMENTS:
Excite
PortSet
Full Bridge (BrFull)
6 Wire Full Bridge (BrFull6w)
Half Bridge (BrHalf)
3 Wire Half Bridge (BrHalf3W)
4 Wire Half Bridge (BrHalf4W)
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
1
3
6
8
2
4
5
7
9060 EXCITATION
C.A.O.
SWITCHED EXCITATION
DIGITAL CONTROL OUTPUT
MADE
IN USA
FIGURE OV1-11. CR9060
CR9070 Counter - Timer / Digital I/O Module — Obsolete
Features 12 channels capable of high-level
(5 V square wave) pulse counting at frequencies up to 5 MHz. Four channels
can also count switch closures; the other eight can count low-level A/C signals.
In addition, there are 16 independent digital I/O channels for digital control,
communications, and triggering.