Section 9. Maintenance — Details
498
supplies approximately 3.6 Vdc. Replace the battery when voltage is
approximately 2.7 Vdc.
•
When the lithium battery is removed (or is allowed to become depleted
below 2.7 Vdc and CR3000 primary power is removed), the CRBasic
program and most settings are maintained, but the following are lost:
o
Run-now and run-on power-up settings.
o
Routing and communication logs (relearned without user
intervention).
o
Time. Clock will need resetting when the battery is replaced.
o
Final-memory data tables.
A replacement lithium battery can be purchased from Campbell Scientific or
another supplier. Table Internal Lithium Battery Specifications
(p. 498)
lists battery
part numbers and key specifications.
Internal Lithium Battery Specifications
Manufacturer
Tadiran
Tadiran Model Number
TL-5902/S
Campbell Scientific, Inc. pn
13519
Voltage
3.6 V
Capacity
1.2 Ah
Self-discharge rate
1%/year @ 20 °C
Operating temperature range
–55 to 85 °C
FIGURE 121: Loosen Thumbscrews
Summary of Contents for CR3000 Micrologger
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ......
Page 30: ......
Page 34: ......
Page 36: ......
Page 96: ......
Page 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Page 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Page 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Page 496: ......
Page 502: ......
Page 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Page 566: ......
Page 594: ......
Page 598: ......
Page 600: ......
Page 602: ......
Page 624: ......
Page 642: ......
Page 643: ......