CS475, CS476, and CS477 Radar Water Level Sensor
•
Distance--the distance between the sensor and water surface. This value
will be reported in either meters or feet, depending on the Units setting.
•
Diagnostic Values—an error code. For example, Code 0 = OK, Code 13 =
error E013 (see Section 7.2).
TABLE 7-1. Example of Start Measurement Command
Initial Command
Response
0M!
Where zero is the
sensor address
00023<cr><If>
Where (from left to right),
0—sensor’s address;
002—the amount of time (in seconds) that you must
wait before sending the send data command;
3—the number of values that will be placed in the
buffer.
Subsequent
Command
Response
0D0!
Where the first zero is
the sensor address
0+100.050+25.000+0<cr><If>
Where (from left to right),
0—sensor’s address;
100.050—the stage in meters or feet;
25.000—the distance in meters or feet
0—error code
7.1.2 Check Unit Response
The Acknowledge Active command is used to check the presence of the sensor
on the bus (see Table 7-2).
TABLE 7-2. Acknowledge Active Command
Initial Command
Response
a!
a<cr><If!>
Only the address is sent back in response.
7.1.3 Check for Valid Data
The aI! command gets the following identification information in response to
sending al! (see Table 7-3).
•
Compatibility level: Version of SDI-12 protocol version. For example, 1.3
•
Manufacturer’s Name: VEGA
17