background image

AVW1/AVW4

3-3

Some additional information concerning the
swept frequency may be found in Appendix C.

NOTE:

  Please remember that sealed (or

absolute) sensors calibrated near sea level
will read negative at higher elevations due
to the decrease in barometric pressure with
increasing elevation.  In most cases the
solution is to load the sensor with extra
external pressure to cause it to operate in
its calibrated range.  For example, a sealed
0 to 5 psi sensor used at 10,000 feet
elevation would have to be placed under
10.6 feet of water (4.6 psi) in order to read 0
psi.  This is because the barometric
pressure at 10,000 feet is about 4.6 psi
lower than at sea level.

3.4  RESOLUTION VS. "# OF CYCLES"

CSI and Geokon recommend that the number
of cycles measured be between 200 and 500.
Five hundred cycles would probably be the
optimum for most applications.  The
measurement of more than 500 cycles is not
recommended without testing.  Some sensors,
for example, will not vibrate a full 1000 cycles at
all pressures in their pressure range.

For the sake of this discussion we will define the
resolution of the vibrating wire measurement as
plus or minus three standard deviations (

±

3 SD)

of the vibrating wire measurement.  The
standard deviation of the measurement in psi is
given by the following equation:

SD = (2*M*C)/(N*T

3

)

3.4-1

where M is the multiplier computed in 3.2-2.  "C"
is 0.00015 ms which is the typical standard
deviation of a time interval measurement on the
CR10 using a 4 mV peak to peak signal with no
external noise.  "N" is the number of cycles
timed to determine the period (T) in ms of the
signal being measured.

The standard deviation is greater when the
period is shorter.  The period is shortest at zero
pressure so the "Period" at zero pressure given
in Table 3.2-1 is used for the example below:

SD = (2*-15.1psi/(kHz

2

) * 0.00015ms)

(500 * (325.6us/1000us/ms)

3

SD = 0.00026246 psi

RESOLUTION = ±

0.00078738

 psi

Resolution improves as the number of cycles
measured increases but it worsens as the
signal level decreases and as the noise
increases.  The optimum resolution is obtained
when the number of cycles measured covers
one time constant (

τ

) of the decay of the signal.

Number of cycles = (

τ

/T)

where T is the period of the signal in
milliseconds and 

τ

 is the time in milliseconds

required for the signal to decay from its initial
peak to peak value to (0.368 * initial peak to
peak value).  An oscilloscope is required to
measure the time constant.  The time constants
of the Geokon sensors we have measured are
typically between one third second and one
second.

The amount of time required to make one
repetition of the vibrating wire measurement is
given below:

150. ms to sweep the frequency

20. ms delay

(1.5 #cycles measured) * period of the
signal in ms

3.5  TEMPERATURE CORRECTION

The equation used to remove errors caused by
changes in the sensor's temperature is found
below:

Pt = P + (TEMP. COEFF.)*(T - To)

3.5-1

where Pt(psi) is the temperature corrected
pressure.  P(psi) is the uncorrected pressure
from equation 1.  T(

o

C) is the current

temperature of the sensor.  To(

o

C) is the

"Temp." on the Calibration Sheet which is the
temperature of the sensor at the time of
calibration.  TEMP. COEFF.(psi/

o

C) is also

found on the Calibration Sheet.  It is important
to note that the "TEMP. COEFF." can be either
positive or negative.

Summary of Contents for AVW1

Page 1: ...UCTION MANUAL Use of the AVW1 and AVW4 with Geokon Model 4500 Vibrating Wire Piezometers and Pressure Transducers Revision 1 92 C o p y r i g h t c 1 9 8 7 1 9 9 2 C a m p b e l l S c i e n t i f i c...

Page 2: ...arranties expressed or implied including warranties of merchantability or fitness for a particular purpose CAMPBELL SCIENTIFIC INC is not liable for special indirect incidental or consequential damage...

Page 3: ...EASUREMENT 3 1 General 3 1 3 2 Multiplier and Offset 3 1 3 3 Swept Frequency Start and End 3 2 3 4 Resolution vs of cycles 3 3 3 5 Temperature Correction 3 3 3 6 Delay Between Measurements 3 4 4 THE A...

Page 4: ...Measurement Error on a 3000 Foot Lead 2 4 2 2 4 Temperature Measurement Error on a 5000 Foot Lead 2 5 2 2 5 Thermistor Linearization Error 2 5 2 3 1 Direct Measurement of the Geokon Thermistor 2 6 3 1...

Page 5: ...l Geokon s manual should be consulted for information on sensor selection and installation Geokon s address and phone number are 48 Spencer Street Lebanon NH 03766 603 448 1562 1 1 SENSOR SELECTION Th...

Page 6: ...some cases there is no measured reference In order to obtain the correct offset under these conditions lower the sensor to a point just above the water level and wait 5 minutes Use the multiplier an o...

Page 7: ...o 60oC The precision of the bridge resistors 0 1 results in a tolerance of 0 03oC The accuracy of the datalogger s voltage measurement 0 015 results in a tolerance of 0 01oC The temperature coefficien...

Page 8: ...5785 0 057855 21 3583 1 304393 21 06127 0 061275 22 3426 1 326119 22 06310 0 063105 23 3277 1 347418 23 06048 0 060484 24 3135 1 368363 24 05747 0 057474 25 3000 1 388888 25 05167 0 051679 26 287 1 40...

Page 9: ...53 965 1 794687 53 09674 0 096746 54 929 6 1 803855 54 08849 0 088499 55 895 8 1 812697 55 07032 0 070322 56 863 3 1 821281 56 04819 0 048193 57 832 2 1 829571 57 01651 0 016519 58 802 3 1 837613 57 9...

Page 10: ...4 FIGURE 2 2 2 Temperature Measurement Error on a 1000 foot Lead Wire is 22 AWG with 16 ohms per 1000 feet FIGURE 2 2 3 Temperature Measurement Error on a 3000 foot Lead Wire is 22 AWG with 16 ohms pe...

Page 11: ...AVW1 AVW4 2 5 FIGURE 2 2 4 Temperature Measurement Error on a 5000 foot Lead Wire is 22 AWG with 16 ohms per 1000 feet FIGURE 2 2 5 Thermistor Linearization Error...

Page 12: ...is manual covers AVW1 s with serial number 1093 and up plus serial numbers 1002 1034 1040A 1041 1042 1051 1052 1055 1057 1058 1059 1069 1071 1073 1076 1080 1084 1086 1087 1088 1088A 1089A 1090A 1091A...

Page 13: ...ry short time The wire will vibrate with the resonant frequency for a relatively long period of time and as it does so it will cut the lines of flux in the plucking and pickup coils inducing the same...

Page 14: ...nd on the Calibration Sheet Example Using sensor number 3998 the multiplier offset and equation for pressure would be M 1000 digits kHz2 0 0151 psi digit M 15 1 psi kHz2 B 0 0 psi 0 0151 psi digit 943...

Page 15: ...les timed to determine the period T in ms of the signal being measured The standard deviation is greater when the period is shorter The period is shortest at zero pressure so the Period at zero pressu...

Page 16: ...required they can be done one of two ways With the AVW4 the repetitions parameter must be set to 2 The instructs the CR10 to excite both sensors delay 20 ms measure the first sensor and then measure...

Page 17: ...nches high The silk screening for the AVW1 is shown below FIGURE 4 1 1 The AVW1 4 2 SENSOR HOOK UP FIGURE 4 2 1 Hook up to AVW1 4 3 WELL MONITORING EXAMPLE In this example the vibrating wire sensor is...

Page 18: ...eet The following is a sample program that measures the temperature and frequency of the vibrating wire sensor and stores the temperature oC pressure psi temperature corrected pressure psi correction...

Page 19: ...P37 Z X F 01 4 X Loc T To C 02 0698 F Temp Coeff 03 4 Z Loc T To C 06 P33 Z X Y 01 2 X Loc PRESS psi 02 4 Y Loc T To C 03 3 Z Loc Pt psi 07 P37 Z X F Converts psi to negative ft H20 01 3 X Loc Pt psi...

Page 20: ...AVW1 AVW4 4 4 This is a blank page...

Page 21: ...wire sensors temperature and pressure to eight single ended CR10 channels The AVW4 has no quiescent current drain The current drain during the very short 2 4 ms temperature measurement is 4 mA per ch...

Page 22: ...ometric pressure only and is used to remove atmospheric pressure changes from the readings on the other three sensors This example assumes the sensors have been connected as shown in the sensor hookup...

Page 23: ...09 0698 B4 06 P87 Beginning of Loop TEMPERATURE CORRECTION 01 00 Delay 02 4 Loop Count 07 P34 Z X F ASSUMES CALIBRATION TEMP 24oC 01 1 X Loc TEMP C 1 02 24 F 03 17 Z Loc T To C 1 08 P36 Z X Y 01 13 X...

Page 24: ...AVW1 AVW4 5 4 This is a blank page...

Page 25: ...A 1 APPENDIX A PRESSURE CONVERSION CHART...

Page 26: ...This is a blank page...

Page 27: ...B 1 APPENDIX B SCHEMATICS AND STUFFING CHARTS FOR AVW1 AND AVW4...

Page 28: ...APPENDIX B SCHEMATICS AND STUFFING CHARTS FOR AVW1 AND AVW4 B 2...

Page 29: ...APPENDIX B SCHEMATICS AND STUFFING CHARTS FOR AVW1 AND AVW4 B 3...

Page 30: ...APPENDIX B SCHEMATICS AND STUFFING CHARTS FOR AVW1 AND AVW4 B 4...

Page 31: ...APPENDIX B SCHEMATICS AND STUFFING CHARTS FOR AVW1 AND AVW4 B 5...

Page 32: ...APPENDIX B SCHEMATICS AND STUFFING CHARTS FOR AVW1 AND AVW4 B 6 This is a blank page...

Page 33: ...How many times do we have to change the half period by 1 clock cycle to cover the frequency range 1 F1 2 1 2400 2 208333 ms half period 1 F2 2 1 3100 2 161290 ms half period Change in half period 2083...

Page 34: ...This is a blank page...

Page 35: ...This is a blank page...

Page 36: ...campbellsci com br suporte campbellsci com br Campbell Scientific Canada Corp CSC 11564 149th Street NW Edmonton Alberta T5M 1W7 CANADA www campbellsci ca dataloggers campbellsci ca Campbell Scientifi...

Reviews: