background image

 

Heating power [W] 

500 

Temperature range [°C] 

Room temperature – 280 

increment: 1 

Temperature display [°C] 

LED 

Temperature display accuracy [°C]  

±1 

The safety temperature of the hotplate [°C] 

320 

Temperature sensor in medium 

PT1000 

Control accuracy of heating temperature 

with temperature sensor [°C] 

±0.5 

Residual heat warning 

50°C 

Dimensions [mm] 

220x160x95 

Weight [kg] 

1.4 

Permitted ambient temperature [°C] 

5-40 

Permitted relative humidity 

80% 

Protection class acc. To DIN 60529 

IP21 

RS232 interface 

Yes 

Accessories  

We have a number of accessories available designed for use with your hotplate 

stirrer. If you need to order any of the below, we invite you to visit our website:  

www.camlab.co.uk

 

Cat No. 

Descriptions 

1189944 

PT1000A Temperature 230mm Probe for all Camlab Choice 

digital hotplates 

1200336 

Support clamp of PT1000 for MS-H280-Pro 

1189932 

Blue carrying plate to hold quarter pies  

1189933 

Blue Fixed Ring  

1189934 

Quarter pie 11 holes 4 ml reaction vessel Ø15.2mm 20mm 

depth  

1189935 

Purple quarter pie 4 holes 20 ml reaction vessel Ø28mm 

24mm depth  

1189936 

Blue quarter pie 4 holes 30 ml reaction vessel Ø28mm 

30mm depth  

1189937 

Black quarter pie 4 holes 40 ml reaction vessel Ø28mm 

43mm depth  

1189938 

Green quarter pie 6 holes 8ml reaction vessel Ø17.8mm 

26mm depth  

1189939 

Golden quarter pie 4 holes 16ml reaction vessel Ø21.6mm 

31.7mm depth  

Summary of Contents for MS-H280-Pro

Page 1: ...MS H280 Pro Round Ceramic Coated Steel Hotplate Stirrer with LED display USER MANUAL 1200464...

Page 2: ...x 6 Permitted conditions of operation 6 Setting up 7 Operation 8 Controls 8 Display 9 Instructions for use 10 Using the temperature sensor 10 Maintenance and Cleaning 11 Faults and Troubleshooting 11...

Page 3: ...Steel Hotplate Stirrer has a double LED display and heats up to 280 C with a stirring speed range of 100 1500rpm Heating and stirring are controlled independently and each function can be used separa...

Page 4: ...pply voltage match and ensure the socket is properly grounded When working with the MS H280 Pro use appropriate PPE to avoid risk of injury caused by o Splashing and evaporating liquids o The release...

Page 5: ...ature sensor is used the tip of the measuring sensor must be at least 5 10mm from vessel bottom and wall The instrument can only be disconnected from the mains power supply by pulling out the connecto...

Page 6: ...pin Bar 50mm x 8mm 1 Permitted conditions of operation This device is designed to operate under the following conditions Indoor use only Use in well ventilated area Permitted ambient temperature range...

Page 7: ...Setting up You must observe a minimum distance of 10cm between the unit and any other object including the wall 10cm Figure 1...

Page 8: ...ependent controls for temperature and speed Both of these functions can be used together and separately if required For the stirring function an appropriate magnetic stirring bar is required Diagram C...

Page 9: ...t 3 Probe Holder Port Port for external temperature probe holder 4 Stir Speed Stirring speed control dial Rotate clockwise to increase speed and counter clockwise to decrease Push button in to turn fu...

Page 10: ...d target temperature and push in the dial to commence heating To turn heating off push the dial in again When heating is not in use the temperature LED will display as OFF 4 Stirring Rotate the dial t...

Page 11: ...ing the cleaning process Only use recommended cleaners below Dyes Isopropyl alcohol Construction materials Water containing surfactant isopropyl alcohol Cosmetics Water containing surfactant isopropyl...

Page 12: ...ines Construction in accordance with the following safety standards EN 61010 1 UL 3101 1 CAN CSA C22 2 1010 1 EN 61010 2 10 Construction in accordance with the following EMC standards EN61326 1 Associ...

Page 13: ...otplate stirrer If you need to order any of the below we invite you to visit our website www camlab co uk Cat No Descriptions 1189944 PT1000A Temperature 230mm Probe for all Camlab Choice digital hotp...

Page 14: ...cts in materials and workmanship under normal use and service for a period of 24 months from the date of invoice and is extended only to the original purchaser It shall not apply to any product or par...

Reviews: