9

When inserting the batteries, ensure correct polarity!

Screw the unit back together carefully without exer-
ting too much pressure on the gasket seal.

If the 

[

!

 

]

 symbol appears together with "bAt” in the

lower section of the display, this means that the batte-
ries are nearly empty and need replacing.

However, the unit will continue to function for a certain
period of time.

If "bAt" appears in the upper section of the display,
this means that the batteries are completely empty.

Only use the matching Camlab oxygen sensors!

Other sensors may cause serious damage to the me-
ter and the sensor.

The sensor or temperature sensor must be plugged in
before the unit is switched on.

Switch the unit on via the [On/Off] key.

The system must be calibrated when the system is
first used.

Following measurement, please switch the unit off
using the [On/Off] key.

See the enclosed Operating Instructions for more de-
tailed information on storage and care of the sensors.

On
Off

Camlab CW/6230  2/5/06  3:33 pm  Page 9

Summary of Contents for CW/6230

Page 1: ...solved Oxygen Meter Instruction Manual Ref CW 72 32 20 Camlab Water Model CW 6230 C O2 hPa mg l mmHg SAL OK Cal hPa MEAS HOLD F C Meas Cal Hold Rcl Store On Off Menu Camlab CW 6230 2 5 06 3 33 pm Page...

Page 2: ...Camlab CW 6230 2 5 06 3 33 pm Page 2...

Page 3: ...unit C F 13 8 7 Selection of unit for oxygen partial pressure 13 8 8 Power off automatic unit switch off 13 9 Oxygen measurement 14 9 1 Absolute pressure 14 9 2 Salinity correction 15 10 Calibration...

Page 4: ...mlab Water CW6230 is a sturdy water tight handheld meter designed for daily measurements in the area of field or laboratory analysis It is used to measure oxygen concentration in surface waters and gr...

Page 5: ...ll them up on the display The standard MEAS measuring function is for conti nuous measurement of oxygen concentrations parti al pressures or saturations The absolute atmospheric pressure is measured i...

Page 6: ...lay 3 Connections C mmHg hPa mg l O2 SAL OK CAL hPa MEAS HOLD F C Meas CAL Hold KCl Store On Off Menu 1 2 3 4 5 6 7 8 9 C pH RealmV mV ATC O CAL MAX MEAS MIN HOLD F C Meas CAL Hold KCl Store On Off Me...

Page 7: ...on O2 Oxygen saturation Secondary display Result temperature or user prompt Percent sensor evaluation C F Temperature unit Warning index battery SAL Salinity correction on off OK Measuring signal stab...

Page 8: ...Calibration mode Store Storage of current result or acceptance of unit configuration Menu Unit configuration 6 Startup The unit is supplied with integrated batteries as stan dard and is therefore rea...

Page 9: ...er section of the display this means that the batteries are completely empty Only use the matching Camlab oxygen sensors Other sensors may cause serious damage to the me ter and the sensor The sensor...

Page 10: ...key takes you into the AUTO HOLD measuring mode In this mode the recorded result is checked by the system in line with the predefined stability criteria until these criteria are met The display shows...

Page 11: ...onfigura tion The unit then switches back to measuring mode You can terminate the configuration process by pres sing the Meas or Hold key Settings made up to this point are not stored The unit retains...

Page 12: ...ter nally during every calibration depending on the sen sor used 100 The electrode is in excellent condition 30 90 The condition of the electrode is ade quate 30 The electrode is old from use or soile...

Page 13: ...F 1 8 C 32 8 7 Selection of unit for oxygen partial pressure hPa Oxygen partial pressure in hPascal mmHg Oxygen partial pressure in mm mercury column 8 8 Power off automatic unit switch off If no key...

Page 14: ...The oxygen partial pressure the oxygen concentrati on mg l and the oxygen saturation are calculated on the basis of the sensor signal and the temperature The measurement is based on steam saturated a...

Page 15: ...Oceanogra phic Tables IOT 10 Calibration of oxygen measurement The sensor data of oxygen sensors are affected by ageing and manufacturing tolerances that result in fluc tuations You should therefore...

Page 16: ...error messages occur during calibration please see the troubleshooting tips If calibration is not completed after you have waited a lengthy period of time at least one of the three measu ring signals...

Page 17: ...lf polarising sensor with silver cathode and lead anode with potassium hydroxide as an elec trolyte is an active system The lead anode the silver cathode and the electrolyte are used up during operati...

Page 18: ...hen disconnecting the sensor do not pull it by the cable but always by the connector sleeve e Remove the sensor storage cap before performing measurements 11 3 Design of the sensor The housing of the...

Page 19: ...maintenance If the sensor is not needed it should be stored either in the storage cap filled with water or directly in a container full of deionised water If a bacterial fungal or algae coating has f...

Page 20: ...disposable syringe Screw the refill screw back in place Fill new membrane cap with electrolyte avoiding air bubbles Screw the sensor vertically into the cap from above When you insert the sensor the...

Page 21: ...he Rcl key the number of the memory location and the corresponding oxygen value are displayed The second time you press the key the temperature for this result is displayed the third time the air pres...

Page 22: ...Cause Remedy Weak battery vol Insert new batteries tage the unit will only continue to function for a short time Batteries empty Insert new batteries No display or Batteries empty Insert new batteries...

Page 23: ...r defective Replace sensor Err 7 System error Send in for repair Err 9 Sensor not present Connect sensor Sensor defective Replace sensor Err 11 Value could not Temperature outside be calculated the ad...

Page 24: ...bubbles in the too high diaphragm head Check calibration environment 14 Technical data Indicating ranges High resolution Low resolution Auto Range Oxygen partial 0 0 570 0 hPa 0 1200 hPa pressure 0 0...

Page 25: ...esult memory locations Power supply 4 x 1 5 V batteries Type AA operating life up to 2000h depending on measuring mode and battery type Power consumption max 0 25 mA depending on measuring mode MEAS o...

Page 26: ...verall length approx 220 mm incl anti kink protection Operating temperature 0 40 C CW 72 32 00 Camlab Water CW6230 Instrument CW 72 32 20 Camlab Water CW6230 Set TT 72 32 01 SensoDirect Oxygen Sensor...

Page 27: ...27 Camlab CW 6230 2 5 06 3 33 pm Page 27...