21
You can enter the temperature of the third buffer solu-
tion using the
[
▲
] [
▼
]
keys.
The value is stored when you press the [STORE] key,
and the status of the electrode is displayed.
N.B.:
A flashing display during manual buffer value input
means that the selected buffer pairs are numerically
too close. The difference between two buffer soluti-
ons must be at least pH 2.00.
10. Result memory
A maximum of 20 results with the corresponding tem-
peratures can be stored.
10.1 Storing results
You can store the current result in the result memory
by pressing the [Store] key.
[Stor] and the numer of the memory location appear
briefly in the display.
10.2 Displaying stored results
You can call the stored results up in the display by
pressing the [RCL] (=Recall) key.
The first time you press the [RCL] key, the number of
the memory location and the corresponding pH or
voltage value (redox value) are displayed.
The second time you press the key, the temperature
for this result is displayed.
Store
Store
+
Rcl
Rcl
Store
Store