35
Scanner
®
2000 microEFM
Section 2
1. Verify that the meter is properly installed in the flow line (per manufacturer’s instructions).
2. Bolt a flange-by-flange 5-valve manifold (as recommended by Cameron) to the Scanner 2000 MVT sen
-
sor.
a. Locate the H and L markings on the integral MVT sensor body and position the MVT/manifold as
-
sembly so that the upstream side of the flow line can easily be connected to the sensor’s “High” port
and the downstream side of the flow line can be connected to the sensor’s “Low” port. The Scanner
2000 enclosure can be rotated to face the desired direction.
b. Position the manifold so that all valves are accessible from the front of the instrument.
3. Connect the Scanner 2000 and manifold assembly to the differential pressure meter. Hardware require
-
ments will vary, depending upon the installation configuration. However, minimally, an adapter is re
-
quired that can span between the threaded pressure tap/orifice flange connector and the non-threaded
manifold. This adapter can be a one-piece stabilizer (often preferred for added strength and stability) or a
short heavy wall pipe nipple attached to a futbol flange (available from Cameron). Use a suitable com
-
pound or tape on all threaded process connections.
CAUTION Do not use Teflon
®
tape on the threads of the union, adapter, or pipe plugs. Use of
Teflon
®
tape will void the explosion-proof rating of the instrument.
4. Install the RTD assembly in the thermowell. Route the RTD assembly cable through the conduit opening
in the top of the Scanner 2000 to connect to the main circuit board. A wiring diagram for the RTD assem
-
Hazardous Area Installations, page 27.
5. Route any additional inputs/outputs or COM connections, etc. through the conduit opening in the top of
the Scanner 2000. For hazardous areas, review
Hazardous Area Installations, page 27.
6.
Perform a manifold leak test as described on
7. Verify the zero offset, if required (and other calibration points, if desired). See the ModWorX™ Pro Soft
-
ware User Manual, Part No. 9A-30165025, for complete instructions. See also
sure or Differential Pressure), page 56, Static Pressure Calibration and Verification, page 56
Pressure Calibration and Verification, page 57
.
CAUTION Do not put the Scanner into operation until the valves are positioned properly so that
pressure is supplied to both sides of the MVT. For instructions on proper valve positions,
see
Placing the Scanner into Operation, page 58
.
Installation Procedure—Remote Mount to Orifice Meter or Cone Meter
A Scanner 2000 can be mounted remotely and connected to an orifice meter or cone meter with tubing for gas
measurement. The setup of the meter run and plumbing configurations can vary widely, depending upon the
challenges existing on location. Figure 2.6 shows a typical remote-mount installation.
WARNING HAZARDOUS AREA USE. The Scanner 2000 is certified for hazardous area use only when
installed in accordance with applicable standards and local wiring practices. Carefully
review
Hazardous Area Installations, page 27,
to determine specific installation require
-
ments (cable glands, conduit seals, signal cable, RTD, etc.).
Summary of Contents for Scanner 2000 microEFM
Page 1: ...NUFLO Scanner 2000 microEFM Hardware User Manual Manual No 9A 30165023 Rev 10 ...
Page 26: ...26 Section 1 Scanner 2000 microEFM ...
Page 104: ...A 18 Appendix A Scanner 2000 microEFM ...
Page 106: ...B 2 Appendix B Scanner 2000 microEFM ...