TLC Sprayer
Instruction manual, May-18
6
2.2
Connect the spray unit
For solutions of regular viscosity, spray head type A is suitable.
For solutions of higher viscosity (such as sulphuric acid up to 30%), CAMAG recommends
to use spray head B (black cover).
•
Remove the cover and the cover disc, screw the spray head (4) with bottle adapter
(7) on the filled reagent bottle (6).
•
Insert the reagent bottle with spray head into the clamp, so that the cam of the
spray head is positioned within the groove of the clamp.
•
Press the end of the connecting tube (8) onto the spray head until it snaps into
position.
By following this procedure, the pump unit is tightly connected with the spray unit.
2.3
Hints
•
Make sure reagent solutions are free from solid particles; filter if necessary.
•
Do not fill the reagent bottle higher than to its shoulder.
•
When using a 50 mL bottle, shorten the capillary tube of the spray head, cutting it
at 45°angle.
•
Do not store the instrument in the charger.
•
Use the sprayer until the battery is almost down, and then fully re-charge it.
•
If the instrument is not used for a longer time, disconnect the mains connection.
3
Getting started
Remove the charged and prepared TLC Sprayer unit from the charger unit and start
spraying by pressing the control switch (9).
CAMAG recommends to start spraying at a distance of 20cm from the plate. Move the
sprayer in a steady motion over the plate until the surface is sufficiently wet (to get best
results, stop before the plate is totally wet).
Spray distance, time and after-spraying treatment depend strongly on the method used.
4
Maintenance/Troubleshooting
4.1
Maintenance
A regular maintenance is mandatory to keep the instrument in perfect working condition.
Therefore follow the steps below:
•
Discharge the sprayer once a month (by spraying until the battery is almost empty)
and charge it fully (also if the TLC sprayer is not used)
•
Replace the spray head on a regular base