Linomat 5
Instruction manual, March 2016
Page 3
16A) outlet. Do not use extension cords without ground con-
tact
•
When working with the fluids of the instrument, be sure to
take the appropriate caution (protect your eyes from direct
contact with liquid)
•
The instrument may be used only by properly trained laboratory
staff
•
The instrument may not be used in rooms with danger of ex-
plosions
•
The instrument contains highly sophisticated electronics and
optical parts. It may be operated only in a non-condensing at-
mosphere in the temperature range outlined in the chapter
"Technical Data". Before installation and use, the instrument
should be acclimated properly
•
Use a damp lint free cloth for cleaning the instrument surface.
Do not employ aggressive detergents
•
Protect yourself and the instrument from electrostatic shock
which can cause damage to the electronic parts
•
Only authorized personnel may open the instrument. Service
and repair is only to be performed by trained specialists. Use
spare parts and consumables supplied by CAMAG only. The
warranty is voided if parts from other sources are used. Check
the service manual before you start service to reduce product-
specific risks
•
The power cord has to be removed before the instrument is
opened. It is not permitted to work on an instrument that has
been opened and is connected to the power supply
•
Spare fuses must be of the type specified by the instrument
manufacturer. It is forbidden to short-circuit or manipulate fus-
es
•
Use only the original, with the instrument delivered power cord
type
•
If the instrument is found to be defective, it must be switched
off and steps must be taken to ensure that it cannot be
switched on by mistake
•
If liquids penetrate the inside of the instrument, the power has
to be disconnected immediately. Small amounts of liquid can
be wiped off and/or dried by means of a hairdryer, with larger
amounts of liquid a service technician has to be called. A test
of functionality has to be performed in all cases
•
Carry out all safety checks and the preventive maintenance as
recommended by the manufacturer in order to assure your per-
sonal safety and the full functionality of the instrument. Have
Summary of Contents for LINOMAT 5
Page 1: ...INSTRUCTION MANUAL LINOMAT 5...
Page 27: ......