User Manual – Rev B
FCS Series II
ENTRY
DESCRIPTION
unit (Master unit). The controller will be disabled
and has no control over the amplifiers, the
measurements or any other function. A message
will be displayed at power indicating it is in Auxiliary
mode. You can press any key to get in the menus
but no control is possible.
LANGUAGE SCPI
APE
The standard bus syntax used by the FCS Series II
is the Standard Commands for Programmable
Instruments (SCPI).
If an alternative syntax suc as APE (Abbreviated
om
h
Plain English) is available, it can be selected fr
this menu.
LAN
If the –LAN option is installed; pressing Enter while
the cursor is on the LANetwork entry provides
access to the LAN interface setting screens listed
below.
IP Address
Displays the IP address setting. This value can be
changed by pressing the SET key and entering a
new value from the keypad or using the Voltage
and Frequency shuttles. Use the numeric data pad
or the voltage shuttle to enter each field. To move
between the four fields, use the decimal point key
he
Host Configuration Protocol (DHCP) mode, press
SET and enter all zeros (0.0.0.0) as the IP address
and cycle power two times. The obtained IP
address will be displayed after the second power
on. For the DHCP setting to work however, the unit
MUST
be connected to a network with a DHCP
server.
Any change to this value will NOT take effect until
after power on the unit has been cycled.
When changing mode from static IP to DHCP, it is
necessary to cycle power on the unit twice, once to
change mode and again to obtain and display a
new IP address from the network.
on the keypad or the Frequency shuttle.
To set a fixed IP address, press SET and enter t
desired IP address. To set the unit to Dynamic
n the
as six
e.g.
’ is
D
MAC Address
Displays the network Media Acces Control
address. This value is fixed and cannot be
changed. The same MAC is normally printed o
model serial tag. The MAC address is shown
hexadecimal numbers separated by a colon,
00:20:4A:9A:02:FD. Note that the leading ‘0
never visible due to the maximum number of LC
characters per line.
1
This feature requires firmware revision 1.40 or higher. If you upgraded from a lower firmware revision, the LAN configuration has
to be enabled to display this menu. Contact customer service for information on enabled this screen.
California Instruments
69
Summary of Contents for FCS Series II
Page 4: ...User Manual Rev B FCS Series II California Instruments 4...
Page 27: ...User Manual Rev B FCS Series II California Instruments 27...
Page 89: ...FCS Series II 89 Figure 5 1 FCS System Block Diagram User Manual Rev B California Instruments...
Page 95: ...User Manual Rev B FCS Series II Figure 6 1 Calibration Setup California Instruments 95...