User and Programming Manual - Rev P
California Instruments
14.2.1. Measurement Subsystem
This subsystem programs the current and voltage measurement capability of the RP Series
AC source.
Subsystem Syntax
:MEASure
:CURRent?
Returns
the
rms
value
of
current
:VOLTage?
Returns the rms value of voltage
MEASure:CURRent?
This query returns the rms value of the output AC current being sourced at the output
terminals. Note that the output relay must be closed to obtain current flow.
Query Syntax
MEASure:CURRent?
Parameters None
Examples MEAS:CURR?
Returned Parameters
<NR2>
Related Commands
OUTP 0
⎜
1
MEASure:VOLTage?
This query returns the AC rms voltage being sourced at the output terminals. Note that the
output relay must be closed or 0 volts will exist at the terminals.
Query Syntax
MEASure:VOLTage?
Parameters None
Examples MEAS:VOLT?
Returned Parameters
<NR2>
Related Commands
OUTP 0
⎜
1
October 2005
RP Series
54