temp The temperature measured inside the power sensor enclosure. The value is in degree Celsius.

sens

The actual input amplifier sensivity. sens is either HIGH or LOW

tflt

The  threshold  fault  state.  if  the  power  reading  is  below  the threshold  the  value  is FAULT,
otherwise it is OK

4.3 The 'set' Command

To  set  the  operational  parameters  the  power  sensor,  request  the  'set' document  with  a  HTTP  GET
command and append the parameters which shall be set to the URL. The path which must be sent with the
GET request is

/set?fmt=txt&...

where  ...  stands  for  an  arbitrary  number  of  parameter  definitions,  each consisting  of  a  key  word  and  the
associated  parameter  value,  separated  by  a '='  character.  Assuming  the  power  sensor  listens  to  the IP
address 10.0.0.1, an example URL for this request is:

http://10.0.0.1/set?fmt=txt&smod=AUTO&offs=0

The  'fmt=txt'  parameter  forces  the  power  sensor  to  reply  a  one  line  text document  rather  than  the  HTML
coded page which is normally displayed by the web browser. The parameter definitions 'smod=AUTO' and
'offs=0' shown in the above example make the power sensor to select it's sensivity range automatically  and
to  clear  any  user  defined  offset  to  the  measurement  value. For  a  complete  list  of  keywords  see  the  table
below.

The power sensor replies to every 'set' request with a complete list of parameters in one line. Like with the
'read' command, the line consists of a set of keyword - value pairs, separated by '&' characters. Within each
pair, keyword and value are separated by the '=' character.

smod=AUTO&fltr=OFF&thrh=-99.99&freq=0&fcor=0.00&offs=0.00&snr=0D8F9

The  values  contained  in  this  list  are  the  values  which  actually  are applying,  including  the  new  values  read
from  the  'set'  URL. The  power  sensor  does  not  give  a  fault  message  if  it  receives  an  invalid parameter
definition.  It  limits  parameter  values  to  a  valid  range  or selections  instead.  If  a  keyword  is  not  known,  it
ignores this parameter completely. To read the parameters actually set with changing a value, send a HTTP
GET request with 'fmt=txt' as the only parameter.

The table below describes the valid parameter settings (keywords and selection tokens are case sensitive):

smod Sets  the  sensivity  range,  the  power  sensor  uses. AUTO  makes  the  power  sensor  select  the

appropriate range  automatically. LOW  selects  the  low  sensivity  range (for  high  levels), HIGH
selects  the  high  sensivity  range. It  is  strongly  recommended  to  operate  the  power  sensor  with
smod=AUTO. The power sensor uses this as fall-back, if it receives an invalid 'smod' selection.

fltr

Sets the filter, the power sensor applies to the measured values. Valid choices are OFFFAST and
SLOWOFF is the default value the power sensor uses if it recognizes an invalid 'fltr' selection.

thrh

Defines the alarm threshold value. If the displayed power value (with any user defines offset and
frequency response calibration already applied) drops below this limit, the power sensor reports a
'low signal' alarm and opens the fault relay contact. To disable the threshold, set it to -99.99.

freq

Defines  the  frequency  of  the  measured  signal  in  MHz.  If this  parameter  is  not  zero,  the  power
sensor  looks  up  the frequency  response  calibration  offset  for  the  given  frequency and  adds  this
offset to each measured value before it is displayed. 'freq' accepts integer numbers on the range 0
.. 19000.

fcor

The  calibration  offset,  the  power  sensor  actually  adds  to  each measured  value  due  to  the

(C) 2020, SatService GmbH

www.satnms.com

PS-UM-2002 Page 11/16

Summary of Contents for SatService sat-nms

Page 1: ...er Manual User Manual Version 1 4 2007 05 07 Copyright SatService Gesellschaft f r Kommunikatiosnsysteme mbH Hardstrasse 9 D 78256 Steisslingen www satnms com www satservicegmbh de Tel 49 7738 99791 1...

Page 2: ...lation 8 3 Operation 8 3 1 The Web based User Interface 8 3 2 Displayed Readings 9 3 3 Operational Parameters 9 4 Remote Control 10 4 1 HTTP Protocol 10 4 2 The read command 11 4 3 The set Command 13...

Page 3: ...C 2020 SatService GmbH www satnms com PS UM 2002 Page 2 16...

Page 4: ...to set the power sensor s IP address which is a essential precondition to operate the power sensor by means of a web browser Operation The sat nms power sensor is operated using a standard web browser...

Page 5: ...ranty Follow standard Electrostatic Discharge ESD procedures when handling an Power Sensor Unit Select and apply the appropriate 24V DC voltage according to the data sheet and documentation before con...

Page 6: ...he power sensor through routers or network switches 2 Now power on your computer and connect the power sensor to the 24V DC supply 3 Insert the CD ROM into the computer s drive and inspect it s conten...

Page 7: ...d Data connectors RF connectors D C and data connectors The RF input connector is a make N type connector mounted at one front panel of the power sensor enclosure At the other side the Ethernet and po...

Page 8: ...s PowerSensor is designed to be mechanically mounted with it s RF input connector For most setups this will be the most convenient method However there are some cases where it is recommended to fix th...

Page 9: ...e you can view and change the operational settings of the power sensor Info After a mouse click to this button the power sensor shows a table with information like the serial number of the device or t...

Page 10: ...values Valid choices are OFF no averaging FAST the sensor displays the average of 8 readings SLOW the sensor displays the average of 48 readings Averaging is done on the base on the 8 or 48 samples th...

Page 11: ...must be sent with the GET request is read fmt txt Assuming the power sensor listens to the IP address 10 0 0 1 the complete URL for the request is http 10 0 0 1 read fmt txt The fmt txt parameter for...

Page 12: ...ually are applying including the new values read from the set URL The power sensor does not give a fault message if it receives an invalid parameter definition It limits parameter values to a valid ra...

Page 13: ...the power which is measured at it s input connector The valid range for offs is 99 99 to 99 99 snr The power sensor s serial number This is a five digit hexadecimal number which uniquely identifies e...

Page 14: ...lies in a satellite ground station or an M C System can be used to provide the supply voltage of the power sensor The 24V supply voltage is routed via a EMV coupled inductor and a transsorber diode wh...

Page 15: ...ns a table where each line consists of a frequency value MHz and the correction value dB separated by a semicolon The file should not contain any other information no space characters and no empty lin...

Page 16: ...t a later time Never change or delete the H TXT L TXT calibration files on the power sensor If these files are damaged the sensor must be returned to SatService GmbH and calibrated again There is no w...

Page 17: ...interface 10 Base T User Interface Via any Web Browser M C Interface for remote monitoring control Via http GET requests Summary fault level alarm indication Photo MOS Relay contact 120 V D C 100 mA E...

Reviews: