12
Address
Length (words)
Description
R/W
0x11
1
Maximum Temperature
R
0x12
1
Filtered Temperature
R
0x13
1
Unfiltered Temperature
R
0x14
1
Sensor Temperature
R
0x15
1
Status (bits active high):
Bit 0: Measurement error
Bit 1: Sensor temperature low
Bit 2: Sensor temperature high
Bit 3: Object temperature low
Bit 4: Object temperature high
R
0x17
1
Reflected Temperature
Not saved to non-volatile memory
Use this register when writing frequently
R/W
0x18
1
Emissivity (1 LSB = 0.0001)
Minimum 0.1000, Maximum 1.0000
Not saved to non-volatile memory
Use this register when writing frequently
R/W
Notes:
1
All temperature are in tenths of degrees C
2
With the exception of addresses 0x17 and 0x18, all write operations are saved to non-
volatile memory
3 For further information please refer to http://www.modbus.org/specs.php
4 The default Modbus slave address of the sensor is 1; it is also possible to use slave
address 255 to communicate. 1 sensor may be connected per COM port.