DEFAULT SETTINGS
The PyroNFC is fully configurable via an iOS / Android app. No configuration is necessary: the
sensor will work straight out of the box using the below default settings.
All settings are adjustable except those marked (fixed).
GUARANTEE
Calex guarantees each instrument it manufactures to be free from defect in material
and workmanship under normal use and service for the period of two years from the date
of purchase. This guarantee extends only to the original buyer according to Calex’s standard
Terms and Conditions of Sale.
Setting
Range of Values
Default Value
(PN151)
Default Value
(PN151K)
Output Voltage*
0-5 V / 1-5 V / 0-10 V 0-10 V
Type K
thermocouple (fixed)
Temperature Low*
0°C to 900°C
0°C
0°C (fixed)
Temperature High*
100°C to 1000°C
1000°C
1000°C (fixed)
Alarm Trigger
High / Low / Off
High
High
Alarm Monitor
Temperature
Target / Internal
Target
Target
Alarm Set Point
0°C to 1000°C
30°C
30°C
Alarm Hysteresis
0°C to 30°C
1°C
1°C
Alarm Mode
Active High / Active
Low
Active High
Active High
Alarm Reset
Auto / Manual
Automatic
Automatic
Emissivity Setting
0.2 to 1.0
0.95
0.95
Reflected Energy
Compensation
On/Off
Off
Off
Reflected
Temperature
0°C to 1000°C
25°C
25°C
Averaging Period
0.125 s to 60 s
0.125 s
0.125 s
Hold Mode
Peak/Valley/Off
Off
Off
Hold Period
0.125 s to 120 s
250 ms
250 ms
* These settings are not adjustable on model PN151K.
Calex Electronics Limited
PO Box 2, Leighton Buzzard, Bedfordshire, England LU7 4AZ
Tel: +44 (0)1525 373178 | [email protected] | www.calex.co.uk
Issue E - Apr 2022
8