Calex PyroNFC Series Operator'S Manual Download Page 3

Mechanical

Construction

Black anodised aluminium and ABS

Cable Length

1 metre standard (longer lengths available to order)

Weight with 1 Metre Cable

65 g

ACCESSORIES

Accessories may be ordered at any time and added on-site. The following accessories are
available:
Fixed mounting bracket - Adjustable mounting bracket - Air purge collar

OPTIONS

Options are factory-installed and cannot be retrofitted. The following options are available:
Certificate of Calibration - Extended cable

INSTALLATION

The installation process consists of the following stages:

Preparation - Mechanical installation - Electrical installation

Please read the following sections thoroughly before proceeding with the installation.

PREPARATION

Ensure that the sensor is positioned so that it is focused on the target only.

AMBIENT TEMPERATURE

The sensor is designed to operate in ambient temperatures from 0°C to 80°C.

Avoid thermal shock. Allow 20 minutes for the unit to adjust to large changes in ambient
temperature.

ATMOSPHERIC QUALITY

Smoke, fumes or dust can contaminate the lens and cause errors in temperature
measurement. In these types of environment, the air purge collar should be used to help keep
the lens clean.

Sensor

CORRECT

INCORRECT

Target larger

than measured
spot

Target smaller

than measured
spot

Background

Distance: Sensor to object (inches)

Distance: Sensor to object (mm)

9

43

76

0.4

1.7

3.0

0

19.7

39.4

0

500

1000

Spot

Diameter

(inches)

(mm)

D:S 15:1

OPTICS

The sensor may be used at longer
distances than shown, and will measure a
larger spot. The measurement accuracy is
not affected by the measurement distance.

There is no maximum measurement
distance provided the air between the
sensor and target is free of obstructions.

DISTANCE AND SPOT SIZE

Mount the sensor at a distance where the
measured spot size is smaller than the
target.

The spot size must not be larger than the
target.

3

Summary of Contents for PyroNFC Series

Page 1: ...PyroNFC Series Operator s Guide...

Page 2: ...Adjustable via app Emissivity Setting Range 0 2 to 1 0 Max Temperature Span Linear Output 1000 C Min Temperature Span Linear Output 100 C Spectral Range 8 14 m Max Supply Voltage 28 V DC Min Supply V...

Page 3: ...perate in ambient temperatures from 0 C to 80 C Avoid thermal shock Allow 20 minutes for the unit to adjust to large changes in ambient temperature ATMOSPHERIC QUALITY Smoke fumes or dust can contamin...

Page 4: ...before switching it back on MECHANICAL INSTALLATION All sensors come with a 1 metre cable and two mounting screws The sensor can be mounted on brackets or cut outs of your own design or you can use t...

Page 5: ...urs and Tags 14 0 13 5 31 0 2 x Hole 3 2 Section A A A A 21 5 THIS FACE AGAINST SENSOR AIR OUT Air Inlet Hose barb 5 mm Touch NFC smartphone here Field of view see OPTICS PWR PWR OP 0 V 21 5 31 0 29 0...

Page 6: ...ntinuously while the phone is touching the sensor On iOS the temperature is updated once each time the user taps update in the app The sensor must be powered to read the measured temperature with the...

Page 7: ...can be solved over the telephone If the sensor is not performing as it should try to match the symptom below to the problem If the table does not help call Calex for further advice TROUBLESHOOTING Sym...

Page 8: ...K thermocouple fixed Temperature Low 0 C to 900 C 0 C 0 C fixed Temperature High 100 C to 1000 C 1000 C 1000 C fixed Alarm Trigger High Low Off High High Alarm Monitor Temperature Target Internal Targ...

Reviews: