INSTALLATION

The installation process consists of the following stages:

 

Preparation 

Mechanical installation 

Electrical installation

Please read the following sections thoroughly before proceeding with the installation.

PREPARATION

Ensure that the sensor is positioned so that it is focused on the target only.

OPTICAL CHART

The optical chart below indicates the nominal target spot diameter at any given distance from the 

sensing head and assumes 90% energy.

1000

103

12

18

M16 x 1mm

11.9

0

Spot Dia. (inches) 

Spot Dia.

(mm) 

Distance: Sensor to object (inches)

Distance: Sensor to object (mm)

61.9

111.9

0.5

2.4

4.4

0

4

8

100

200

D:S 2:1

Distance: Sensor to object (inches)

Distance: Sensor to object (mm)

11.9

36.9

61.9

0.5

1.5

2.5

0

19.7

39.4

0

500

1000

Spot Dia. (inches) 

Spot Dia.

(mm) 

D:S 20:1

1

PCAN21

PCAN201

Sensor

CORRECT

INCORRECT

Background

Target greater 

than spot size

Target smaller

than spot size

[email protected]

www.digiparts.ch

Ihr Schweizer Industriepartner

Summary of Contents for PyroCAN Series

Page 1: ...PyroCAN Series Operators Guide info digiparts ch www digiparts ch Ihr Schweizer Industriepartner...

Page 2: ...x 103 mm long Thread Mounting M16 x 1 mm pitch Cable Length 1 m Weight with Cable 95 g ENVIRONMENTAL Environmental Rating IP65 Ambient Temperature 0 C to 90 C Relative Humidity 95 max non condensing A...

Page 3: ...spot diameter at any given distance from the sensing head and assumes 90 energy 1000 103 12 18 M16 x 1mm 11 9 0 Spot Dia inches Spot Dia mm Distance Sensor to object inches Distance Sensor to object...

Page 4: ...AL INTERFERENCE To minimise electromagnetic interference or noise the sensor should be mounted away from motors generators and such like WIRING Check the distance between the sensor and the connected...

Page 5: ...Water flow rate should not be more than 0 5 to 1 5 litres min 40 0 21 19 63 103 M16 x 1mm 1 8 BSP water air connections 1 8 BSP Air connection 25 APSW for PCAN21 50 APSN for PCAN201 20 29 40 AIR PURGE...

Page 6: ...e ID is persistent between power cycles The CAN ID may be set to a value from 0 to 2048 0x0 to 0x800 as a 4 byte unsigned integer The emissivity setting may be set to a value from 0 2 to 1 0 as a 4 by...

Page 7: ...x00 0x00 0x00 Command ID Send the command with these bytes empty The sensor will return the emissivity setting in these bytes as a floating point value After the following commands are sent the sensor...

Page 8: ...lex guarantees each instrument it manufactures to be free from defect in material and workmanship under normal use and service for the period of one year from the date of purchase This guarantee exten...

Reviews: