Dokument
IMA 190
HFE507
Rev
5
Sid
36 (44)
Part of the service menu for setting the MAX and MIN values.
6a
AOUt
yes
First step of adjusting the analogue outputs MAX and MIN values
6b
An xx
no
Select analogue output by stepping with MENU(+) button
6c1
SetL
no
Leads to adjustment of the MIN value.
6d1
The present
MIN value in
% of FS
no
Increase the MIN value by stepping with the MENU(+) button.
Decrease by stepping with the ESC(-) button. The output is set to
the MIN value and can be checked with a multimeter.
6e1 Sure
no
The adjustment is saved by pressing Entr(+-). Press ESC to return
without saving.
6c2
SetH
no
Leads to adjustment of the MAX value
6d2
The present
MAX value in
% of FS
no
Increase MAX value by stepping with the MENU(+) button. Decrease
by stepping with the ESC(-) button. The output is set to the MAX
value and can be checked with a multimeter.
6e2 Sure
no
The adjustment is saved by pressing Entr(+-). Press ESC to return
without saving.
PLEASE NOTE!
If a power failure happens when the sensor has OUT1...4 locked to
min/max limits, then the sensor will have this output locked when the power returns.
It is necessary to enter this menu item and release the output manually!