background image

s.Cod.

= Access code for calibration menu.

This parameter should be

.

See NOTE 1 for programming.

222

A.CAL.

= Current calibration.

At this state, 20.000 mA current is applied to the
input of the device.

10.CA.

= 10V input calibration.

At this state, 10.000V is applied to the input of
the device.

1.CA.

= 1V input calibration.

At this state, 1.0000V is applied to the input of
the device.

For initializing calibration, first

then

keys are pressed together and held until ‘

'

message appears. See NOTE 3.

CAL

To initialize calibration, first

then

keys are pressed together and held until ‘

'

message appears. See NOTE 3.

CAL

To initialize calibration, first

then

keys are pressed together and held until ‘

'

message appears. See NOTE 3.

CAL

If increment key

is pressed and held 0.6 seconds, the value of the selected parameter changes rapidly. If

waited enough, the value increases 100 at each step. After 1 second following the release of the key, initial
condition is returned. The same procedure is valid for the decrement key.

For including decimal point first,

then,

keys are pressed and held together. And

then, by using

key decimal point can be

adjusted.

S.Cod.

d.C.sc.

Uo.sc.

D.Ca.s.

S.Cod.

a.CAL.

1.CA.

10.CA.

If

key is pressed and held for 5

seconds programming mode is entered

U.opt.

d.Cal.

SeCU.

d.pnt.

Cal.t.

l.sCl.

H.sCl.

l.inp.

H.inp.

CAL.t.=U.ýnP.

Evet

No

d.Cnf.

i.typ.

rate

Hold

dsp.C.

Unit

If first

and then

keys are

pressed together , measurement
unit appears. See
For programming.

Unit

parameter

If first

and then

keys are

pressed together, the maximum and the
minumum measurement values become
equal to the measured value at that time

and the message

appears on

display.

res.

Pressing this key the minumum
measured value up to that time
appears.

Pressing this key the maximum
measured value up to that time
appears

The minumum
measured value

The maximum
measured value

Measurement unit

870

1453

Bar.

&

Programming Mode

3/3

Running Mode

Measurement
value

1000

Res.

-E-05

EI141

SET

SET

SET

SET

SET

SET

SET

CAL.t.

= Calibration type.

Selectable as

or

.

If

is selected, input type is one the

four standard input types. If

is

selected, input types can be modified.
See NOTE 1 for modification.

S.inP

S.inp

U.inP

.

.

U.inP

i.typ.

= Input type.

Input type can be selected as 0-20mA,
4-20mA, 0-1V or 0-10V.
See NOTE 1 for programming.

RAtE

= Sampling rate.

Measurement is performed at each 200ms.
However,
for

rate, each measurement is displayed.

FASt

for

rate, the average of 4 successive

measurements is displayed.

SLo.1

for

rate,

for

rate,

SLo.2

SLo.3

the average of 8 successive

measurements is displayed.

the average of 16 successive

measurements is displayed.
See NOTE 1 for programming.

Hold

= Display holding parameter.

Selecting

, this parameter becomes

inactive.
Selecting

, always the minimum

measured value is displayed.

NonE

Lo.

Hi.

,

Selecting

always the maximum

measured value is displayed.

dSP.C.

= Display configuration.

Selectable as

or

. If

is

selected, process value appears. If
is selected, process value and then
measurement unit are displayed 4 and 2
seconds successively.

PrcS.

Pr.Un.

PrcS.

Pr.Un.

See NOTE 1 for programming.

Unit

= Measurement unit.

A constant, a message etc. to be displayed
can be entered. If a decimal point is desired,
it should be included before entering the
character.

d.Pnt.

= Decimal point.

Decimal point can be adjusted between
1. and 3. digits.
See NOTE 1 for programming.

L.SCL.

= Lower limit for the scale.

It can be adjusted between -1999 and
(

-100).

H.SCL

.

See NOTE 1 for programming.

H.SCL.

= Upper limit for scale.

It can be adjusted between (

+100)

and 4000.

L.SCL.

See NOTE 1 for programming.

SET

At this state, the reference voltage or
current that corresponds to

Parameter

is applied to the input.

To initialize calibration, first

then

keys are pressed together and held until ‘

'

message appears. See NOTE 2.

L.SCL.

CAL

SET

At this state, the reference voltage or
current that corresponds to
parameter is applied to the input.

To initialize calibration, first

then

keys are pressed together and held until

' message appears. See NOTE 2.

H.SCL.

CAL

SET

SET

SET

s.Cod.

= Access code for safety menu.

This parameter should be
See NOTE 1 for programming.

333

.

d.C.Sc.

=

menu protection level

parameter.

= No menu is seen.

= Menu is seen but can not be programmed.
= Menu is seen and programming is possible.

ConF.

nonE
P. no
P.yES

.

See NOTE 1 for programming.

Uo.Sc.

=

U.oPt.

menu protection level

parameter.

= No menu is seen.
= Menu is seen but can not be programmed.

= Menu is seen and programming is possible.

See NOTE 1 for programming.

nonE
P. no
P.yES

.

d.Ca.S.

=

d.CAL.

menu protection level

parameter.

= No menu is seen.
= Menu is seen but can not be programmed.

= Menu is seen and programming is possible.

See NOTE 1 for programming.

nonE
P. no
P.yES.

CAL.

CAL.

C.end

C.end

s.err.

S.err.

C.err.

C.err.

The message on the left flashes approximately 5
seconds and calibration is completed.

The message on the left flashes approximately 5
seconds and calibration is completed.

If calibration is error free, the message

appears

. However,

if it is wrong, the message
appears for 1 seconds and the program
is shifted to the next step.

C.End

S.Err

.

for 1 seconds

If calibration is error free, the message

appears

. However,

if it is wrong, the message
appears for 1 seconds and the program
is shifted to the next step.

C.End

C.Err.

for 1 seconds

If the difference between the reference voltages or currents applied for the calibration of

and

is lower than one

half of the full scale, this error message appears on the display.
For example: Assume that the selected input type is 0-1V. In this case, if the difference

voltages

applied for calibration of

and

is lower than 0.5V, this error message appears.

H.inP.

L.inP.

H.inP.

L.inP.

between the reference

If the reference voltage or current applied to the input for calibration is too high or too low, this error message appears.

ERROR MESSAGES

CAL okey?

CAL okey?

CAL okey?

Yes

Yes

No

No

NOTE 2

NOTE 3

SET

For adjusting a selected parameter first press and hold

key. Then, by using

keys adjustment

can be made.

Parameter adjustment method

NOTE 1

Entering from programming mode to run mode:

If no key is pressed within 20 seconds during programming mode data is stored automatically and the run
mode is entered. Alternatively the same function occurs by pressing

key and holding for 5 seconds

SET