Stroller/Sprint
10
Warning:
Do not operate Stroller/Sprint while
attached to the stationary unit.
Warning:
Should the leakage be excessive to the
point that a stream of liquid is present, leave the area
and call your health care provider immediately.
Caution:
If the Stroller/Sprint is being refilled
immediately after a period of use, fill it, and then
wait 10 minutes after filling with the FCV set at zero
for the pressure to stabilize. Then operate as
normal.
Warning:
Liquid or cold gaseous oxygen can frostbite
the skin. Oxygen will result in increased fire hazard.
Portable
Release
Button
(push-on
style only)
Closing the valve at the first sign of liquid discharge
does not fill the Stroller/Sprint as completely, but saves
liquid oxygen.
9. For the push-on style connectors, push the release
button down until the units separate. For all other
connectors rotate the unit until it separates from the
stationary unit.
Caution:
Check liquid level gauge only after the vent
valve is closed.
Caution:
If the Stroller/Sprint does not separate easily,
do not use force. The units may be frozen together.
Wait until the units warm up — then they’ll separate
easily. Do not touch any frosted parts. Should there be
any liquid leakage from the stationary unit or
Stroller/Sprint after separation, reattach the
Stroller/Sprint immediately. Again separate the units. If
the problem persists, leave units coupled and call your
health care provider.