MN240-1 Rev C
37
5.13
Hour Meter Replacement
HM009-2 is a dual function hour meter. It has a selection switch accessible by a pinhole in
the upper right corner of the hour meter.
To toggle between display modes - press in the switch once using a small paperclip.
1.
‘TMR1’ mode displays accumulated hours, which can be reset to zero hours by pressing
and holding the switch in until 0.0 is displayed (approximately 3 seconds).
2.
If ‘TMR1’ is not displayed, then total operating hours of unit is displayed.
1.
Set the unit’s ON/OFF switch to the OFF position and disconnect the power cord.
2.
Remove the left side panel.
3.
Cut the heat shrink to expose the hour meter’s leads.
4.
Disconnect the hour meter’s leads.
5.
Push hour meter mounting tabs inward from the hour meter.
6.
Remove the hour meter through the control panel.
7.
Install the new hour meter into the control panel, making sure the hours are
displayed on top. (Hour glass icon should be located on bottom of hour meter when
properly installed).
8.
Reconnect the hour meter leads.
9.
Reconnect the side panel.
5.14
Flowmeter Replacement
Flowmeter Removal
To remove the flowmeter, follow the steps below:
1.
Set the unit’s ON/OFF switch to the OFF position and disconnect the power
cord.
2.
Remove the side panel.
3.
Cut the tie-wraps and remove the ¼-inch green oxygen tubing from the
flowmeter fittings.
4.
Unscrew the flowmeter nuts with a wrench.
5.
Remove the flowmeter through the control panel.
Flowmeter Installation
To install a new flowmeter, follow the flowmeter removal procedure in reverse
order. Leak test the tubing connections.
Summary of Contents for NewLife Elite
Page 1: ...MN240 1 Rev C 1 NewLife Family Technical Manual MN240 1...
Page 6: ...MN240 1 Rev C 6...
Page 7: ...MN240 1 Rev C 7...
Page 8: ...MN240 1 Rev C 8...
Page 9: ...MN240 1 Rev C 9...
Page 10: ...MN240 1 Rev C 10...
Page 11: ...MN240 1 Rev C 11...
Page 46: ...MN240 1 Rev C 46...
Page 47: ...MN240 1 Rev C 47...
Page 48: ...MN240 1 Rev C 48...