MN240-1 Rev C
34
5.9
Product Regulator Check and Setting
The product regulator is factory-set at 7 PSIG (48kPa) for the Elite, 20 PSIG (138 kPa) for
the Intensity
8 and
10 and should not require adjustment. To check for proper adjustment
of the product regulator, take the following steps:
Procedure A: NewLife Elite
1.
Set the ON/OFF switch to the ON position.
2.
Allow the unit to run for five minutes.
3.
Turn the flowmeter adjustment knob counter-clockwise until it stops (wide
open).
4.
The flowmeter ball centers itself on the 5.5 LPM line. If not, the product
regulator needs to be adjusted.
Procedure B: NewLife Intensity
8 and
10
1.
Install outlet pressure test kit (TU131-1) to the outlet of the NewLife unit.
2.
Set the unit’s ON/OFF power switch to the ON position.
3.
Allow the unit to run for five minutes.
4.
Turn the flowmeter adjustment knob counter-clockwise until it stops (wide
open).
5.
The outlet pressure should measure 20 PSIG for the Intensity 10. If the outlet
pressure is not at 20 PSIG the product regulator needs to be adjusted.
5.9.1
Setting Product Regulator for Normal Operation
Procedure A: NewLife Elite
Use the following procedure to set the product regulator to 7 PSIG (48kPa):
1.
Disconnect the humidifier bottle, if used, and the tubing from the oxygen outlet.
2.
Plug in the unit.
3.
Set the unit’s ON/OFF switch to the ON position, and allow the unit to run at least
five minutes to build up pressure.
4.
Remove the right-side panel.
5.
Turn the flowmeter adjustment knob counterclockwise until it stops (wide open).
6.
Pull outward on the regulator knob to unlock it.
7.
Turn the regulator knob or set screws until the flowmeter ball centers on the 5.5
LPM line (clockwise to increase).
8.
Push in the regulator knob to lock it.
9.
Reconnect the side panel.
Summary of Contents for NewLife Elite
Page 1: ...MN240 1 Rev C 1 NewLife Family Technical Manual MN240 1...
Page 6: ...MN240 1 Rev C 6...
Page 7: ...MN240 1 Rev C 7...
Page 8: ...MN240 1 Rev C 8...
Page 9: ...MN240 1 Rev C 9...
Page 10: ...MN240 1 Rev C 10...
Page 11: ...MN240 1 Rev C 11...
Page 46: ...MN240 1 Rev C 46...
Page 47: ...MN240 1 Rev C 47...
Page 48: ...MN240 1 Rev C 48...