background image

21

Lib
er

at

ors

Troubleshooting Charts

XI

Symptom

Probable Cause

Corrective Action

7)

(Cont.)

Low  flow  at  oxygen 

outlet  on  all  LPM 

settings (cont.)

d)

Flow 

  

   

control 

  

   

valve

  

 

   

inlet 

         

filter 

  

screen 

  

dirty

Clean or replace (RPxx) the filter screen.

e)

Economizer valve faulty

Test the economizer (RP28) and replace (RP29) if necessary.

f )

Blockage

      

   

 in

   

 

    

the

   

 

  

liquid 

  

   

 

    

withdrawal

 

circuit or the flow restrictor

Check

   

 the 

   

warming 

   

coils 

   

and 

   

withdrawal 

   

tubes

   

 as 

   

well 

   

as 

   

the 

   

flo

 

w

       

 restrictor 

        

for 

blockages.  Replace if necessary.

g)

FCV faulty

Replace the FCV (RP30)

8)

Increased NER

a)

Saturation

 

 

 

            

Pressure

          

 

          

    

is

                           

too

        

 

         

high

Inspect 

 

  

the saturation pressure of the reservoir.  Allow at least 30 minutes at 

no flow for the

 

 

 

reservoir to saturate properly. 

b)

Leak in the system

Perform a leak check on the plumbing (RP16).  Repair leaks as necessary.

c)

Relief valve open

Ensure that there is no venting from the relief valves.  If there is refer to the 

corrective actions for “Excessive venting from relief valves (hissing sound)”

.

d)

Partial 

  

or complete los

s

 

of

     

 

   

vacuum

Conduct 

    

the

 

 NER 

  

test (RP32) and 

  

return 

  

the

 

 unit 

 

to CAIRE, Inc.

 if

   

 necessary.

9)

  

Excessive      Frost:

NOTE: Minimal    frost 

on   the shroud   and on 

the   plumbing      

is normal. This 

symptom      applies    

to  frost that is 

much greater  than   

what    is   normally    

observed.

a)

Frost is acceptable

Some 

  

frost 

 

on

 

 the 

 

shroud

  

 and on the plumbing is acceptable, 

  

especially 

   

at 

  

high 

  

flow 

rates  during  continuous  use.    This  is  due  to  the  evaporation  of  LOX  to  gas  and  the 

temperature 

        

difference

 

between the LOX and room temperature.

b)

High humidity level

High humidity levels can increase frost accumulation.

c)

Saturation                

     pressure       is   too   

h igh

Inspect

       

 the saturation pressure of the reservoir.  Allow

at least 30 minutes at no flow for the reservoir to saturate properly. 

d)

Leak in the system

Perform 

  

   

leak 

   

check 

   

on 

    

the

    

 plumbing (RP16).  

   

Repair 

   

leaks 

   

as

   

 necessary.

e)

Relief valve open

Ensure that there is no venting from the relief valves.  If there is refer to the 

corrective actions for “Excessive venting from relief valves (hissing sound)”

f )

Partial                               or                            complete                                 loss                                  of                                                         vacuum

Conduct the NER test (RP32) and return the unit to CAIRE, Inc. if necessary.

10)

Unit  will  not  maintain 

acceptable 

system 

pressure

a)

Saturation pressure is 

unacceptable

Inspec

t    the    saturation    pressure    of      

 the

       

 reservoir

        

 used

 for          filling  . 

.  Allo

w   at   

lea

st 30       

 

 

minut

es   at  

no flow for the portable to saturate properly. 

b)

Vent                                        valve                                                  not                                                 completely                                               closed

Close   vent   valve. Leak   check    vent    valve    outlet    and    stem.   Replace   or    repair    as    needed.

c)

Leak in the system

Perform a leak check on the plumbing (RP16).  Repair leaks as necessary.

d)

Economizer valve faulty

Test the economizer (RP28) and replace (RP29) if necessary.

e)

Primary releif valve faulty

Test the primary relief valve (RP13) and replace (RP14) if necessary

11)

High  Pressure  at 

Reservoir  oxygen 

outlet

a)

Saturation

   

 pressure 

   

is 

   

too

    

 high Inspect the saturation pressure of the reservoir.  Allow at least 30 minutes at 

no flow for the reservoir to saturate properly. 

b)

Primary  reli

e

f  valve  set  too  high  or 

relief 

    

valve

     

 operating

     

 improperly.

Perform relief valve test. Replace as needed.

c)

Economizer

      valve        stuck        in 

closed         state

Perform Economizer test. Replace as needed.

d)

Parti

 

al

                       

 

   

or

      

 complete

           

 lo

ss    of      vacuu

m

Conduct

  

 the 

   

NER 

   

test (RP32) and 

   

return

   

 the 

   

unit

   

 to 

    

CAIRE, Inc. 

    

if 

    

necessary.

12)

Contents

      

 indicator 

reads incorrectly

a)

Battery

      

 voltage

      

 to

o low    to

 

operate

     

 contents indicator

Replace 

       

battery 

             

as

           

 needed

    

 (RP5) 

          

then

           

calibrate

          

 the

              

 meter

              

 following

         

 (RP6)

b)

Sensor wires not connected properly, 

pinched or otherwise damaged

Visually inspect the meter harness assembly from the meter to the point 

where the sensor wire enters the reservoir manifold. Replace as necessary

.

c)

Contents

      

 indicato

r     not   

calibrated 

           

properly

Calibrate the meter as needed following (RP6).

d)

Ice crystal in reservoir causing 

incorrect or empty level reading

Empty and warm the reservoir following (RP23, RP24) to melt and evaporate 

moisture from inside the reservoir.

e)

Meter

            

 malfunctioning/

damaged

Inspect for physical damage. Replace meter as necessary then recalibrate the 

meter follow-ing (RP6)

Liberators Technical Service Manual • PN 13350704 Rev 

K

 

Summary of Contents for LIB20

Page 1: ...Liberators G4 Technical Service Manual...

Page 2: ...ic Rim Phone 770 721 7759 Fax 770 721 7758 Europe Phone 44 0 1189 367060 Fax 44 118 9799245 For Customer or Technical Service needs product assistance ordering part numbers specifications unexpected e...

Page 3: ...ption of a condition that can result in equipment or component damage NOTE A statement containing information important enough to emphasize or repeat ITEM Item numbers used throughout this manual are...

Page 4: ...ry of Operation 11 VII Unpacking and Setup 14 VIII Operation 15 IX Maintenance 17 X Troubleshooting 19 XI Troubleshooting Charts 20 23 XII Repair Procedures 24 XIII Service Tools Equipment Supplies 38...

Page 5: ...tempera ture and regulates the flow of oxygen gas to the patient Any water that condenses on the cold coils is routed into the con densate bottle An economizer regulator is utilized to con serve LOX...

Page 6: ...bar 23 2 psi 1 6 bar 23 2 psi Primary Relief Valve Reseat Pressure 1 4 bar 20 5 psi 1 4 bar 20 5 psi 1 4 bar 20 5 psi 1 4 bar 20 5 psi 1 4 bar 20 5 psi Secondary Relief Valve Opening Pressure 2 1 bar...

Page 7: ...as Association as an oxygen concentration above 23 In an oxygen enriched atmosphere flammable items may burn vigorously and may ex plode Certain items considered non combustible in air may burn rapidl...

Page 8: ...s are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B Harmonic emissions The Liberator is suitable for use in all establishments incl...

Page 9: ...input output lines No data input output lines 1 kV line s Surge to line s Not Applicable Not Applicable IEC 61000 4 5 2 kV line s DC powered device to earth Voltage dips 5 UT 95 dip short interruptio...

Page 10: ...nce level in each frequency rangeb Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These gui...

Page 11: ...2 3 P 0 01 0 12 m 0 12 m 0 23 m 0 1 0 38 m 0 38 m 0 73 m 1 1 2 m 1 2 m 2 3 m 10 3 8 m 3 8 m 7 3 m 100 12 m 12 m 23 m For transmitters rated at a maximum output power not listed above the recommended...

Page 12: ...e is substantially higher than the Liberator operating pressure oversaturated 2 Saturation pressure of the liquid oxygen in the fill source is substantially lower than the Liberator operating pressure...

Page 13: ...the bypass tee to the breathing coil the flow control valve and then the patient As the pressure in the container increases over 1 4 bar 20 psig the economizer valve opens and the cycle repeats mainta...

Page 14: ...e meter is powered by an internal battery offering battery life of 5 years or more at 30 cycles per day The meter has a low battery LOW BATT indicator which signals the need for bat tery replacement T...

Page 15: ...vel meter by depressing the operate button LEDs will light displaying the level of liquid oxygen If the unit is empty only the leftmost LED should light If it does not or if other erroneous indication...

Page 16: ...transfer line adapter to transfer line d Fully open liquid valve on fill source e Purge transfer line for a minimum of 5 seconds ensuring gas is safely piped away from operator i Connect transfer hose...

Page 17: ...tongs or tweezers 2 Wipe off the exterior of the unit with the cleaning solution us ing a lint free cloth The cloth should be damp but not drip ping Be as thorough as possible 3 Discard the cleaning c...

Page 18: ...nual NOTE See the Troubleshooting and Repair section of the cor responding technical manual on the detailed procedures for the tests mentioned below 1 Visual Inspection a Remove any LOX prior to maint...

Page 19: ...cator functionality d Presence of all required labels e Cryogenic reservoir damage dents dings f If LOX is still present in the unit inspect for heavy frost or condensation on the exterior of the unit...

Page 20: ...mbly RR 31 RP18 Vent Valve RR 31 RP19 Twist Lock QDV Lip Seal RR 31 RP20 Twist Lock QDV RR 32 RP21 Push on QDV RR 32 RP22 Push on Portable Pop Off Assembly 32 RP23 Empty Unit 33 RP24 Warm Unit 33 RP25...

Page 21: ...e QDVs are wiped clean and dry before filling b Dirty or damaged QDV poppet Replace the QDV cartridge RP20 or the entire QDV assembly RP21 5 Excessive venting from relief valves hissing sound a Satura...

Page 22: ...ionsfor Excessive venting from relief valves hissing sound f Partial or complete loss of vacuum ConducttheNERtest RP32 andreturntheunittoCAIRE Inc ifnecessary 10 Unit will not maintain acceptable syst...

Page 23: ...22 Liberators XI Troubleshooting Charts 8 Liberators Technical Service Manual PN 13350704 Rev K...

Page 24: ...e Troubleshooting Chart Start at the upper left corner The top line shows the steps of routine maintenance Unless otherwise noted by the arrows the flow through the chart is down or to the right Liber...

Page 25: ...attery WARNING Make sure the unit is empty and vent valve is open before replacing any component except shroud assembly compo nents or Lip Seals NOTE When assembling new compression fittings tighten 1...

Page 26: ...n more accurate but re quires the dewar be vent full with LOX To start either procedure calibration mode must first be entered There is no need to per form both procedures for any given calibration Pr...

Page 27: ...el 20 30 37 45 60 LED Min Max Min Max Min Max Min Max Min Max 1 17 6 20 6 22 0 24 7 22 7 27 9 24 9 31 3 34 2 38 1 2 20 6 23 5 24 7 28 9 27 9 33 1 31 3 37 6 38 1 46 3 3 23 5 26 5 28 9 33 2 33 1 38 3 37...

Page 28: ...7 Range The new full cap value is less than empty cap value When calibrating full Empty unit and recalibrate TABLE 6 Calibration Errors TABLE 7 Operation Errors LEDs Name Reason Notes Solution 2 4 6 H...

Page 29: ...e three bolts Item 31 that run through the FCV bracket Item 39 and through the manifold plate into the top of the dewar g Remove two remaining manifold mounting bolts Item 31 not visible in Figure 12...

Page 30: ...l holding signal wire against the ca pacitance probe c Unsolder feed thru wire Item A from probe Item B d Loosen feed thru nut Item C and remove harness e To replace manifold harness reverse above pro...

Page 31: ...Item 28 should open audible venting and or bubbling of leak test solution i Decrease pressure to 1 3 bar 19 psig Test relief valve with leak test solution A minimal amount of leakage bubbling is acce...

Page 32: ...o humidifier adapter Item 26 on FCV outlet Open FCV Item 24 to 6 LPM setting d Assemble oxygen regulator pneumatic hose Item C with DISS fittings and male pneumatic test adapter Item D Connect as semb...

Page 33: ...em 88 on tool f Install retaining ring Item 87 in QDV body Item15 by turning tool counter clockwise while pushing in on tool Torque retaining ring Item 87 to 225 280 N cm 20 25 in lbs RP17 Warming and...

Page 34: ...e static seal Item 67 if needed FIGURE 21 Top Fill Push on QDV Assembly WARNING Liberator must be empty and vented before starting procedure WARNING Liberator must be empty and vented before starting...

Page 35: ...g 4 Set Liberator FCV Item 24 to 6 LPM setting 5 Continue until Liberator is empty RP24 Warm Unit Figure 23 a Liberator may be allowed to sit FCV off vent closed a mini mum of 48 hours after emptying...

Page 36: ...alues within tolerances listed in Table 9 or unit fails flow rate test NOTE Flow rates will only be accurate if the reservoir is filled properly Reference the filling section of this manual for the co...

Page 37: ...ected to normal temperature and pressure 70 F 14 7 psig or 21 C and 1 01 Bar The above acceptable flow rates are intended for field testing only of flow control valves to account for variation in elev...

Page 38: ...ator Item 35 c Follow RP29 to ensure proper settings on the replacement regulator d Reverse remainder of procedure to install new properly ad justed regulator RP29 Economizer Regulator Test a Follow s...

Page 39: ...en in a full Liberator 45 reservoir 1 Empty weight of LIB 45 Reservoir 24 95 kg 55 00 lbs 2 Filled unit with properly saturated liquid oxygen 3 Full weight of LIB 45 Reservoir 74 99 kg 165 32 lbs 4 74...

Page 40: ...uper Flex Liquid Oxygen Transfer Line 6 97406630 Dual Fill Head Tee 13350704 Service Manual 10661515 Conversion Kit TF to SF 10660361 Conversion Kit SF to Dual 10660344 Conversion Kit TF to Dual 10661...

Page 41: ...40 Liberators Parts Price List XIV Contact Customer Service or visit www cairemedical com to obtain your parts list Liberators Technical Service Manual PN 13350704 Rev K...

Page 42: ...Fax 44 0 1344 429224 4 Mail or fax the completed purchase order for confirmation to CAIRE Inc 2200 Airport Industrial Dr Ste 500 Ball Ground GA 30115 USA CAIRE Medical Ltd Unit 6 Ashville Way Wokingha...

Page 43: ...ntenance performed Restocking Policy If it becomes necessary to cancel an order with CAIRE Inc af ter the shipment has been received use the following Restock Policy procedure 1 Notify the Customer Se...

Page 44: ...Wokingham Berkshire RG41 2PL United Kingdom Ph 44 0 1189 367060 Fax 44 0 0118 979 9245 Copyright CAIRE Inc 2019 CAIRE Inc reserves the right to discontinue its products or change the prices materials...

Reviews: