CAIRE Eclipse 5 User Manual Download Page 6

6 - ENG    PN 21182453 C | User Manual

Eclipse 5

Your unit is equipped with an Air Inlet Filter (extra 

filter provided). Check daily and clean once a week.

Connect the oxygen supply 

tubing to the oxygen outlet and 

connect the oxygen cannula 

per the instructions provided 

with the cannula, or per the 

instructions from your home 

care provider.

Breathe normally through the nasal cannula.

When operating on battery power, utilizing the 

Pulse Dose Mode will achieve longer dura-

tion-of-use.

7 Universal Cart Operation

Align the locater holes on the bottom of the device 

to the tabs on the Universal Cart. Align the large 

threaded screw to the insert at the back of the 

Eclipse 5. Then hand-tighten the knob on the cart 

until the Universal Cart is tightly secured to the 

device. Depress the push-button on the telescoping 

Universal Cart handle to adjust the height of the 

handle.

6 Begin Using the Eclipse 5

Air Inlet Filter

Pulse Dose Mode Display 

4 Select Flow Delivery Mode

Flow Mode Button and Indicator

Press the Flow Mode button to select your desired 

flow mode. Pressing this button repeatedly will 

toggle you back and forth between continuous 

flow and pulse dose modes. In Continuous Flow 

Mode the oxygen is provided at a constant flow rate 

between 0.5 and 3.0 LPM. In Pulse Dose Mode, ox-

ygen is supplied in a bolus at the beginning of each 

inspiration, providing a selectable range of settings 

1-9 with bolus sizes 16-192 mL. Pulse Dose can be 

adjusted for sensitivity and breath bolus delivery. 

Flow settings in both modes should be prescribed 

for you by your physician.

5 Adjust Flow Setting to Prescribed 
Level

Increase or Decrease Flow Setting 
Button

Using the + or - buttons, adjust the setting to the 

prescription supplied by your physician.

Continuous Flow Mode Display

WARNING: IT IS VERY IMPORTANT TO 
SELECT ONLY THE PRESCRIBED LEVEL OF 
OXYGEN. DO NOT CHANGE THE FLOW SE-
LECTION UNLESS YOU HAVE BEEN DIRECT-
ED TO DO SO BY A LICENSED CLINICIAN. 
THE PORTABLE OXYGEN CONCENTRATOR 
MAY BE USED DURING SLEEP UNDER 
THE RECOMMENDATION OF A QUALIFIED 
CLINICIAN. YOUR DELIVERY SETTINGS 
SHOULD BE PERIODICALLY REASSESSED 
FOR EFFECTIVENESS OF THERAPY.

Contact your equipment provider for available 

training options. If you do not know your provider, 

contact CAIRE.

Note: For assistance, if needed, in setting up, or 
maintaining the Eclipse 5, or to report unexpect-
ed operation or events, contact CAIRE or CAIRE 
Representative.

Summary of Contents for Eclipse 5

Page 1: ...Eclipse 5 with autoSAT User Manual English...

Page 2: ...ial Number Reg 2498 This way up Reg 0623 Fragile handle with care Reg 0621 Keep away from rain keep dry Reg 0626 Stacking limit by number Reg 2403 ISO 7010 Graphical symbols Safety colors and safety s...

Page 3: ...sieve This material separates the oxygen from the nitrogen and allows only the oxygen to pass through The result is a flow of high concentration oxygen delivered to the user The Eclipse 5 combines ad...

Page 4: ...al Become familiar with the key features of the Eclipse 5 and the User Control Panel Eclipse 5 AC Power Supply DC Power Supply Universal Cart Power Cartridge battery FRONT Control Panel Handle Oxygen...

Page 5: ...nd Indicator The button toggles between Contin uous Flow and Pulse Dose Mode The Pulse Dose Mode activates autoSAT Technology and allows a significant increase in the operating time while powered by t...

Page 6: ...back and forth between continuous flow and pulse dose modes In Continuous Flow Mode the oxygen is provided at a constant flow rate between 0 5 and 3 0 LPM In Pulse Dose Mode ox ygen is supplied in a...

Page 7: ...ION ABOUT THE DISCOMFORT AND OR THE MEDICAL URGENCY TO THE RESPONSIBLE CARE GIVER TO AVOID HARM PATIENTS WITH HEARING AND OR SIGHT IMPAIRMENT S MAY NEED ASSISTANCE WITH MONITORING ALARMS Contraindicat...

Page 8: ...ENTRATOR THE BATTERY MUST ALSO BE REMOVED FROM THE DEVICE AND PACKAGED PROPERLY WARNING DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTAND ING THIS MANUAL IF YOU ARE UNABLE TO UNDERSTA...

Page 9: ...ING THE US DEPARTMENT OF TRANS PORTATION DOT AND UNITED NATIONS UN REGULATIONS REQUIRE THE REMOVAL OF THE BATTERY FROM THE DEVICE FOR ALL INTERNA TIONAL AIRLINE TRAVEL WHEN THE OXYGEN CONCENTRATOR IS...

Page 10: ...IF UNABLE TO LEAVE THE ROOM YOU MUST WAIT 10 MINUTES AFTER YOU HAVE TURNED THE OXYGEN CONCENTRATOR OFF BEFORE SMOKING WARNING NO SMOKING OXYGEN IN USE SIGNS MUST BE PROMINENTLY DISPLAYED IN THE HOME O...

Page 11: ...llow these instructions can result in the power supply not supplying power to the Oxygen Concentrator CAUTION When the automobile in which you are using the Oxygen Concentrator unit is turned off disc...

Page 12: ...t Nominal Battery Recharge Time 1 8 to 5 0 hrs recharge time to achieve 80 capacity dependent upon the flow Alarms Alerts Loss of power Low Battery Low Therapeutic O2 Output O2 flow outside normal lim...

Page 13: ...et filter at the top rear of the cabi net and allow exhaust air to freely leave the exhaust vent at the bottom left of the device Place the device such that the alarms can be heard Position the oxygen...

Page 14: ...326 AC POWER SUPPLY WITH TWO PINS The Eclipse 5 includes a universal AC Power Supply for use at home or wherever standard AC power is available To connect to an AC power source be sure the AC Power Su...

Page 15: ...le from the vehicle s DC power outlet WARNING DO NOT ALLOW EITHER THE AIR INTAKE OR THE AIR OUTLET VENTS TO BECOME BLOCKED THIS CAN CAUSE THE OXYGEN CONCEN TRATOR TO OVERHEAT AND IMPAIR PERFORMANCE WA...

Page 16: ...RES OR WITH A FULLY CHARGED OR COMPLETELY DISCHARGED BATTERY CAN DEGRADE ITS OVERALL BATTERY LIFE DO NOT ATTEMPT TO OPEN THE BATTERY THERE ARE NO SERVICEABLE PARTS INSIDE THE BATTERY KEEP BATTERIES AW...

Page 17: ...Note Operating the Oxygen Concentrator outside of its normal operating temperature range can affect performance and decrease battery run time and or increase battery charge time Refer to the Specific...

Page 18: ...ot detected The device will continually check for an inspiratory effort every 15 seconds Once an inspiration is detected the device reverts to measured bolus delivery To exit Pulse Dose Mode and engag...

Page 19: ...7 feet 2 1 m of tubing PN 5408 SEQ For any additional recommended accessories please see the Accessories Catalog PN ML LOX0010 available on www CAIREmedical com A firebreak is required for use with an...

Page 20: ...he yellow light is on Battery symbol is flashing on LCD The Eclipse 5 power cartridge voltage may be low 1 Plug Eclipse 5 into AC power supply and recharge or replace power cartridge with fully charge...

Page 21: ...battery CHARGER ON OR OFF ON 1 alarm beep repeat every 20 seconds 001 Ambient pressure fail AMB PRESS ON OR OFF ON 1 alarm beep repeat every 20 seconds 004 O2 85 After warm up O2 LOW OFF ON 1 alarm be...

Page 22: ...sformer as well 3 Ensure that the power verification lights are on There will be a green light on the power supply box There will be a green light on the Eclipse 5 concentrator control panel in the sh...

Page 23: ...be a green light on the Eclipse 5 concentrator control panel in the shape of a power cord 4 If power verification LEDs are not lit remove all connections of the power supply for 20 seconds and reconne...

Page 24: ...r in warm tap water using a mild soap detergent solution 3 Rinse the filter thoroughly with clear tap water and squeeze out the excess water 4 Allow the filter to air dry 5 Reinsert the filter in the...

Page 25: ...BASED PROD UCTS CONCENTRATED CHLORINE BASED PROD UCTS ETHYLENE CHLORIDE AND OIL BASED PRODUCTS PINE SOL LESTOIL THESE ARE NOT TO BE USED TO CLEAN THE PLASTIC HOUS ING ON OXYGEN CONCENTRATOR AS THEY CA...

Page 26: ...ssions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The device is suitable for use in all establishments including domestic es...

Page 27: ...mmended that the Eclipse 5 is powered from an uninterruptible power supply UPS Power frequency magnet ic field IEC 61000 4 8 30A m Power frequency magnetic fields should be at levels characteristic of...

Page 28: ...28 ENG PN 21182453 C User Manual Eclipse 5 NOTES...

Page 29: ...PN 2118245 C User Manual 29 ENG Eclipse 5 NOTES...

Page 30: ...30 ENG PN 21182453 C User Manual Eclipse 5 NOTES...

Page 31: ...2 3 3 4 5 5 6 7 8 Table is based on 20 Breaths Per Minute for Pulse Dose Mode only 7 1 2 2 3 3 4 5 5 6 6 8 1 2 3 3 4 5 6 6 7 8 9 1 2 3 4 5 6 7 8 8 9 Table is based on 15 Breaths Per Minute for Pulse...

Page 32: ...2018 Chart Industries Chart Industries reserves the right to discontinue its products or change the prices materials equipment quality descriptions specifications and or processes to its products at a...

Reviews: