PN 2118245 C | User Manual 21 - ENG
Eclipse 5
Alarms Table
Alarm
LCD Text
Green
Yellow
Audio
Code
All OK, warm-up done
NO ALARM
ON
OFF
OFF
000
Warm-up O2 < 85%
O2 LOW
ON
ON
No Beeps
008
No breath detected (less than
45 seconds)
NO ALARM
ON
OFF
OFF
000
No breath detected (exceeding
45 seconds)
P <--> C
ON
ON
1 alarm beep, repeat
every 20 seconds
200
One wire communication loss
ONE WIRE
ON OR
OFF
ON
1 alarm beep, repeat
every 20 seconds
100
Cannot charge battery
CHARGER
ON OR
OFF
ON
1 alarm beep, repeat
every 20 seconds
001
Ambient pressure fail
AMB PRESS
ON OR
OFF
ON
1 alarm beep, repeat
every 20 seconds
004
O2 < 85% (After warm-up)
O2 LOW
OFF
ON
1 alarm beep, repeat
every 20 seconds
008
Flow rate error > +/-10%
FLOWRATE
OFF
ON
1 alarm beep, repeat
every 20 seconds
020
Warm battery > 60°C
WARM BAT
ON
ON
1 alarm beep, repeat
every 20 seconds
002
Low battery
(5% remaining battery
capacity)
LOW BAT
ON
ON
1 alarm beep, repeat
every 20 seconds
010
Loss of power
********
OFF
ON
1 alarm beep, repeat
every 20 seconds for
200 seconds
040
Malfunction
FAIL XX
OFF
ON
1 alarm beep, repeat
every 20 seconds for
200 seconds
080